41
20901224_EN_VACUU·PURE 10C_V1.8_221122
Installation and connection
Connect the outlet line
1
2
(3)
(1)
(2)
(4)
(4)
1.
1. Connect rubber sealing ring (3)
(3)
, hose nozzle
(2)
(2)
, and union
nut
(1)
(1)
as shown and screw them onto the outlet connection.
2.
2.
Push outlet line
(4)
(4) onto the hose nozzle and, if necessary,
lay the hose in a fume hood. Fix the outlet line, e.g., with a
hose clip.
>
The length of the outlet line at the vapor condenser must
not exceed 3 m. An excessively long outlet line can lead to
inadmissibly high back pressure at the outlet and impair the
function of the sealing gas. Use an outlet line with an interior
diameter of at least 19 mm.