67
20901224_EN_VACUU·PURE 10C_V1.8_221122
Error messages
Error
Error
`
Possible cause
3
Remedy
Personnel
No or
minimal
suction
power.
`
Leak in inlet line or in
apparatus.
3
Check inlet line
and equipment for
leaks.
Operator
`
Optional separator
not correctly fitted
or sealing ring on
separator missing.
3
Check and correctly
install separator.
`
Inlet line too long
or cross‑section too
small.
3
Use shorter inlet
line with larger
cross‑section.
`
Condensate inside
the vacuum pump.
3
Let vacuum pump
run for a few
minutes with suction
nozzle open or in
regeneration mode.
`
Deposits inside the
vacuum pump.
3
Rinse vacuum
pump.
Specialist
`
High level of vapor
generated in the
process.
3
Check process
parameter.
`
Pump speed reduced
due to excessive
temperature.
3
See warning,
flashing rate 1x.
Resp.
specialist
Key LEDs do
not light up.
`
Vacuum pump
switched off.
3
Switch on vacuum
pump at toggle
switch.
Operator
`
Power plug not
correctly plugged in
or pulled out.
3
Check power supply
and cable.
`
Vacuum pump
defective.
3
Send in vacuum
pump.
Resp.
specialist