63
20901559_EN_RE_RZ_V7_230523
Appendix
Vacuum data
Max. pumping speed 50 / 60 Hz
(US)
Rx 2.5
2.3 / 2.8 m
3
/h
1.4 / 1.7 cfm
Rx 6
5.7 / 6.8 m
3
/h
3.4 / 4.0 cfm
Ultimate partial pressure without gas ballast
RE 2.5
3*10
‑1
mbar
2.3*10
‑1
Torr
RZ 2.5
4*10
‑4
mbar
3*10
‑4
Torr
RE 6
1*10
‑1
mbar
7.5*10
‑2
Torr
RZ 6
4*10
‑4
mbar
3*10
‑4
Torr
Ultimate total pressure without gas ballast
RE 2.5
3*10
‑1
mbar
2.3*10
‑1
Torr
RZ 2.5
2*10
‑3
mbar
1.5*10
‑3
Torr
RE 6
1*10
‑1
mbar
7.5*10
‑2
Torr
RZ 6
2*10
‑3
mbar
1.5*10
‑3
Torr
Ultimate total pressure with gas ballast
RE 2.5
8*10
‑1
mbar
6*10
‑1
Torr
RZ 2.5
1*10
‑2
mbar
7.5*10
‑3
Torr
RE 6
6*10
‑1
mbar
4.5*10
‑1
Torr
RZ 6
1*10
‑2
mbar
7.5*10
‑3
Torr
Max. permissible inlet pres‑
sure, absolute
1.1 bar
16 psi
Max. permissible outlet pres‑
sure, absolute
1.1 bar
16 psi
Max. permissible pressure
difference between inlet and
outlet
1.1 bar
16 psi
Max. permissible pressure at
gas ballast valve, absolute
1.2 bar
17.5 psi
Mechanical data
Dimensions (L x W x H), approx.
(US)
Rx 2.5
(230 V)
389 mm x 156 mm
x 240 mm
15.3 in x 6.1 in
x 9.4 in
Rx 2.5
(100 – 120 V / 220 – 230 V)
389 mm x 156 mm
x 240 mm
15.3 in x 6.1 in
x 9.4 in
Rx 6
(230 V)
420 mm x 156 mm
x 240 mm
16.5 in x 6.1 in
x 9.4 in
Rx 6
(100 – 120 V / 220 – 230 V)
420 mm x 156 mm
x 240 mm
16.5 in x 6.1 in
x 9.4 in
Technical data
Summary of Contents for RE 2.5
Page 24: ...24 20901559_EN_RE_RZ_V7_230523 Introduction ...
Page 40: ...40 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Page 48: ...48 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Page 73: ...73 20901559_EN_RE_RZ_V7_230523 Appendix 8 8 8 8 CU Certificate CU Certificate CU Certificate ...
Page 76: ...76 20901559_EN_RE_RZ_V7_230523 ...
Page 77: ...77 20901559_EN_RE_RZ_V7_230523 ...
Page 78: ...78 20901559_EN_RE_RZ_V7_230523 ...