17
20901559_EN_RE_RZ_V7_230523
Safety information
DANGER
DANGER
Occurrence of explosive mixtures in the
vacuum pump or at the outlet.
Mechanically generated sparks, hot surfaces or static
electricity may ignite explosive mixtures.
>
Prevent any formation of explosive mixtures in the
expansion chamber, in the oil reservoir, or at the out‑
let of the pump.
>
Connect inert gas for gas ballast or venting.
>
Drain appropriately or otherwise remove any poten
‑
tially explosive mixtures at the outlet of the pump.
>
Dilute any potentially explosive mixtures at the outlet
of the pump to non‑explosive concentrations.
>
Comply with applicable regulations when disposing of chemi
‑
cals. Take into consideration that chemicals may be contami
‑
nated.
>
Wear your personal protective equipment and take precau
‑
tions to avoid contact with the skin, inhalation, and possible
irritations.
Hazards due to different substances
Pumping different substances or media can cause the substances
to react with one another.
>
Be aware of interactions and possible chemical reactions of
the pumped media.
>
Purge the vacuum pump with ambient air or inert gas before
changing the medium to be pumped. Thereby you will pump
residues out of the vacuum pump and reduce the possibili‑
ty of reactions of the pumped substances with each other and
with the pump’s materials.
Pumping different
substances
Summary of Contents for RE 2.5
Page 24: ...24 20901559_EN_RE_RZ_V7_230523 Introduction ...
Page 40: ...40 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Page 48: ...48 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Page 73: ...73 20901559_EN_RE_RZ_V7_230523 Appendix 8 8 8 8 CU Certificate CU Certificate CU Certificate ...
Page 76: ...76 20901559_EN_RE_RZ_V7_230523 ...
Page 77: ...77 20901559_EN_RE_RZ_V7_230523 ...
Page 78: ...78 20901559_EN_RE_RZ_V7_230523 ...