23
20901559_EN_RE_RZ_V7_230523
Safety information
2.5
2.5 Motor protection
Motor protection
CAUTION
Impaired lock of the thermal cutout in case of
supply voltage below 115 VAC.
In case of supply voltage below 115 V, the lock of the
thermal cutout might be impaired. The pump may re
‑
start on its own after sufficient cooling down.
>
In case of excess temperature, switch off the vacuum
pump or disconnect the pump from the power source
to avoid an automatic restart.
The pump motor is equipped with a self‑hold thermal cutout in the
winding as overload protection. In case of excess temperature the
vacuum pump switches off.
If the vacuum pump is switched off due to these safe
‑
ty precautions, the error must be cleared manually:
Unplug vacuum pump from power supply
eliminate cause of
error
switch vacuum pump back on after a waiting time of 5 min
‑
utes. Allow the vacuum pump to cool down prior to switching on
again.
2.6
2.6 Disposal
Disposal
NOTE
NOTE
Electronic devices must not be disposed of in do-
mestic waste at the end of their service life.
Used electronic devices contain harmful substances that
can cause damage to the environment or human health.
Disused electrical devices also contain valuable raw materials,
which can be recovered for reuse if the device is disposed of
correctly within the recycling process.
End users are legally obliged to take used electric and
electronic devices to a licensed collection point.
>
Correctly dispose of all electronic scrap and electronic com
‑
ponents at the end of their service life.
>
Observe the national regulations regarding
disposal and
environmental protection.
Overheating
protection
Summary of Contents for RE 2.5
Page 24: ...24 20901559_EN_RE_RZ_V7_230523 Introduction ...
Page 40: ...40 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Page 48: ...48 20901559_EN_RE_RZ_V7_230523 Installation and connection ...
Page 73: ...73 20901559_EN_RE_RZ_V7_230523 Appendix 8 8 8 8 CU Certificate CU Certificate CU Certificate ...
Page 76: ...76 20901559_EN_RE_RZ_V7_230523 ...
Page 77: ...77 20901559_EN_RE_RZ_V7_230523 ...
Page 78: ...78 20901559_EN_RE_RZ_V7_230523 ...