48
NivoGuide® 8100 • Two-wire 4 … 20 mA/HART
7 Diagnostics and servicing
58878-EN-190215
Error
Cause
Rectification
4 … 20 mA signal missing
•
Electrical connection faulty
•
Check connection according to chapter
"
Connection steps
" and if necessary,
correct according to chapter "
Wiring
plan
"
•
Voltage supply missing
•
Check cables for breaks; repair if
necessary
•
Operating voltage too low or load
resistance too high
•
Check, adapt if necessary
Current signal greater than
22 mA or less than 3.6 mA
•
Electronics module in the sensor
defective
•
Exchange the instrument or send it in
for repair
The below tables show typical examples for application-relevant
measurement errors. There are two measurement errors:
•
Constant level
•
Filling
•
Emptying
The images in column "
Error pattern
" show the real level as a broken
line and the level displayed by the sensor as a continuous line.
1
2
Level
time
0
Fig. 19: The broken line 1 shows the real level, the continuous line 2 shows the
level displayed by the sensor
Note:
•
Wherever the sensor displays a constant value, the reason could
also be the fault setting of the current output to "
Hold value
"
•
If the level indication is too low, the reason could be a line resist-
ance that is too high
Measurement error with constant level
Fault description
Cause
Rectification
1. Measured value shows a
too low or too high level
Level
time
0
•
Min./max. adjustment not correct
•
Adapt min./max. adjustment
•
Incorrect linearisation curve
•
Adapt linearisation curve
•
Running time error (small measure-
ment error close to 100 %/serious
error close to 0 %)
•
Repeat setup
Treatment of measure-
ment errors
Summary of Contents for NivoGuide 8100
Page 72: ...72 NivoGuide 8100 Two wire 4 20 mA HART INDEX 58878 EN 190215 U Units 26 ...
Page 73: ...73 NivoGuide 8100 Two wire 4 20 mA HART Notes 58878 EN 190215 ...
Page 74: ...74 NivoGuide 8100 Two wire 4 20 mA HART Notes 58878 EN 190215 ...
Page 75: ...75 NivoGuide 8100 Two wire 4 20 mA HART Notes 58878 EN 190215 ...