NivoGuide® 8100 • Two-wire 4 … 20 mA/HART
6 Set up with the display and adjustment module
58878-EN-190215
•
Special parameters
The copied data are permanently saved in an EEPROM memory in
the display and adjustment module and remain there even in case of
power failure. From there, they can be written into one or more sen-
sors or kept as backup for a possible electronics exchange.
Note:
Before the data are stored in the sensor, a check is carried out to
determine if the data fit the sensor. If the data do not fit, a fault signal
is triggered or the function is blocked. When data are being written
into the sensor, the display shows which instrument type the data
originate from and which TAG-no. this sensor had.
Tip:
We recommend to save the instrument adjustments. In case of an
electronics exchange the saved parameter adjustment data relieve
this process.
Since scaling is very extensive, scaling of the level value was divided
into two menu items.
In menu item "
Scaling variable
" you define the scaling variable and
the scaling unit for the level value on the display, e.g. volume in l.
In menu item "
Scaling format
" you define the scaling format on the
display and the scaling of the measured level value for 0 % and
100 %.
Since scaling is very extensive, scaling of the interface value was
divided into two menu items.
Additional settings - Scal-
ing level
Additional settings -
Scaling level - Scaling
variable
Additional settings - Scal-
ing level - Scaling format
Additional settings - Scal-
ing interface
Summary of Contents for NivoGuide 8100
Page 72: ...72 NivoGuide 8100 Two wire 4 20 mA HART INDEX 58878 EN 190215 U Units 26 ...
Page 73: ...73 NivoGuide 8100 Two wire 4 20 mA HART Notes 58878 EN 190215 ...
Page 74: ...74 NivoGuide 8100 Two wire 4 20 mA HART Notes 58878 EN 190215 ...
Page 75: ...75 NivoGuide 8100 Two wire 4 20 mA HART Notes 58878 EN 190215 ...