11.4
KEYPAD
parameter menu
Parameters such as limit values,
mA output or, if option "controller"
has been selected, controller
parameters are set in the P menu.
function menu
Settings such as zero point,
measured value and measuring
range can be set in the F menu.
They are used to select values in
direction keys
the P or F menus and for the
manual operation of controller
outputs.
key MANUAL/
Activates manual operation mode
AUTOMATIC
required for selection of values in
the P or F menus.
key "E"
If settings in the P or F menus are
changed (e.g. limit value increased
from 0.30 to 0.35 mg/l), the
B-18.PCX - edition 01/1993
selected value has to be entered
by pressing key "E".
11.5
PARAMETER AND FUNCTION MENUS -
SET-UP AND CHANGING OF SETTINGS
1)
Press key
"Hd." is displayed in the text display,
MANUAL/AUTOMATIC:
manual operation mode of ampflifier is
activated.
2)
Press keys "P" or "F":
The respective menus are displayed in
the text display.
Press key again to scroll menu.
3)
Press direction keys:
Values to be changed can be selected.
4)
Press key "E":
The selected value is entered.
5)
Press key
Return from F or P menus to manual
MANUAL/AUTOMATIC:
operation mode.
34
Summary of Contents for DEPOLOX 4
Page 10: ......
Page 41: ......
Page 54: ......
Page 55: ......
Page 56: ......
Page 60: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......
Page 108: ......
Page 109: ......
Page 110: ......
Page 111: ......
Page 112: ......
Page 113: ......