6.- EQUIPMENT MAINTENANCE
- 75 -
6.2.- Care of cables
The UM2B unit is capable of producing high voltages, as a result of
which THE CABLES SHOULD BE IN PERFECT ORDER, TO
AVOID THE DANGER OF ELECTRICAL SHOCK OR ERRORS
AND INACCURACIES IN MEASUREMENT.
The cables and their condition should be periodically checked in order to detect
beforehand any deterioration or breakage that might cause situations of danger for the
operators and/or malfunctioning of the equipment. If the cables are damaged, they should be
sent for repair to an authorized technical service or new cables should be acquired (see
chapter 8). This is applicable also to the cable and unit connectors.
Special care should be taken with the serial cable for connection of the PC, since it is
in charge of controlling the unit.
It is very important not to tread on or move the cables during testing,
since this might alter the measurements, especially at low current values
(very high measured resistances).
For properly measuring, the cables must be completely stretched until the
measurement point without creases, or anyway, that these have a radius
not inferior than 200 mm.
Summary of Contents for UM2B
Page 2: ......
Page 67: ...5 SOFTWARE DESCRIPTION 65 Figure 5 46 a First page of the report Recovery Voltage ...
Page 68: ...5 SOFTWARE DESCRIPTION 66 Figure 5 47 a Second page of report Recovery Voltage ...
Page 69: ...5 SOFTWARE DESCRIPTION 67 Figure 5 48 Third page of report Recovery Voltage ...
Page 70: ...5 SOFTWARE DESCRIPTION 68 Figure 5 46 b First page of the report Insulation Resistance ...
Page 71: ...5 SOFTWARE DESCRIPTION 69 Figure 5 47 b Second page of the report Insulation Resistance ...
Page 87: ...8 TECHNICAL SUPPORT 85 Description of fault Materials returned on of of the year Table 8 1 ...
Page 95: ...APPENDIX B CONTROL SOFTWARE 93 APPENDIX A CE CONFORMITY DECLARATION ...