background image

UTG9000T Series Function Abitrary Waveform Generator                                          

 

31 

Tap Coupling to step through AC/DC. The default option is AC. 

3)  Trigger Level 

Tap TrigLeve

pop-out the virtual numeric keyboard to enter numbers and tap return 

key.The default value is 0V. 

4)  Sensitivity 

Tap sensitivity, pop-out the virtual numeric keyboard to enter numbers and tap return 
key.The default value is 100%. 

5)    High Frequency Reject 

Tap HFReject to step through OFF/ON. The default option is OFF 

6)    Clear 

Tap Clear to delete all the measured result and restart to calculate.   

 

2.4.5 Internet Settings 

Table 2-7 Internet Settings   

Menu 

Sub-menu 

Setting 

Description 

Internet Setting 

DHCP 

OFF/ON 

Turn off to set the following option 

IP address 

 

 

Mask 

 

 

Gateway 

 

 

MAC 

 

 

 
Tap Utility

System to enter the Internet settings interface (on the right side) 

1.  DHCP 

Tap DHCP to step through OFF/ON.   

2.  IP Address 
  IP address format: nnn.nnn.nnn.nnn, the first of nnn range can set to 1~233, the second 0~255. 
It is suggest that you inquiry the internet administrator to get an available IP address. Tap IP or 
push the numeric keyboard to enter numbers. This settings will be stored in non-volatile memory, 
and the instrument will automatically apllied on next boot. 
3.  Mask 

Subnet mask format: nnn.nnn.nnn.nnn, nnn range can set to 0~255. It is suggest that you 

inquiry the internet administrator to get an available subnet mask address. Tap Mask or push the 
numeric keyboard to enter numbers. This settings will be stored in non-volatile memory, and the 
instrument will automatically apllied on next boot.   
4.  Gateway 

Summary of Contents for UTG9000T Series

Page 1: ...User s Manual UTG9000T Series Function Arbitrary Waveform Generator ...

Page 2: ...sing this brand new product In order to use this product safely and correctly please read this manual thoroughly especially the safety notes After reading this manual it is recommended to keep the manual at an easily accessible place preferably close to the device for future reference ...

Page 3: ... new products after repair All replaced components modules and products will be properties of UNI T The customers below are individuals or entities of rights provided in the warranty according to the statement In order to obtain services promised in the warranty customers must report defects to UNI T during the applicable warranty period and make appropriate arrangement for performance of services...

Page 4: ... distributors refuse to give any implied warranty for marketability or applicability for special purpose For violation of the warranty repair or replacement of defective products is the only and all remedial measure UNI T provides for customers No matter whether UNI T and its distributors are informed of any possible indirect special occasional or inevitable damage in advance they assume no respon...

Page 5: ...Sa s the vertical resolution 16bits Adjustable noise bandwidth Sine wave output 600MHz 400MHz 200MHz full band 1μHz Square wave output 200MHz 160MHz 120MHz the minimum edge time within 1 5ns adjustable duty ratio Pulse wave output 200MHz 160MHz 120MHz wide dynamic range high precise adjustable rising falling edge time adjustable duty ratio It can output phase and amplititude independent and adjust...

Page 6: ...me Hardware frequency counter 800MHz AC DC current coupling Powerful upper computer software and arbitrary editor 10 1 capacitive touch screen 1280 800 resolution Standard configration interface USB Host USB Device LAN independent input and output 10MHz colock source Easy to use multi purpose knob and numeric keyboard Offer NeptuneLab system management software ...

Page 7: ... wave 19 2 3 1 Frequency Output Settings 19 2 3 2 Output Amplitude Settings 20 2 3 3 DC Offset Voltage Settings 20 2 3 4 Squre Wave Settings 21 2 3 5 Pulse Wave Settings 21 2 3 6 DC Voltage Settings 22 2 3 7 Ramp Wave Settings 22 2 3 8 Noise Wave Settings 23 2 3 9 Hamonic Wave Settings 23 2 3 10 PRBS Wave Settings 24 2 3 11 Noise Superposition Settings 25 2 4 Secondary Function Settings 25 2 4 1 C...

Page 8: ...ger source 109 3 2 6 Trigger Output 110 3 2 7 Trigger Edge 111 3 2 8 Comperhensive example 111 3 3 Output Burst Waveform 113 3 3 1 Select Burst 113 3 3 2 Burst Type 115 3 3 3 Initial Phase of Burst 116 3 3 4 Burst Period 117 3 3 5 Counting of Burst 117 3 3 6 Select Trigger Source 117 3 3 7 Trigger Output 118 3 3 8 Trigger Edge 118 3 3 9 Comprehensive example 119 3 4 Output Abitrary Waveform 121 3 ...

Page 9: ...000T Series Function Abitrary Waveform Generator 9 5 3 Contact US 143 Appendix A Factory Setting 145 Appendix B Performance Index 149 Appendix C Accessories List 159 Appendix D Maintenance and Cleaning 160 ...

Page 10: ...t may immediately happen WARNING Indicates an injury or hazard that may not immediately happen CAUTION Indicates that a potential damage to the instrument or other property might occur Symbols on the product The following symbols may used on the product AC Measuring ground terminal Frame ground terminal ON OFF Danger High voltage Caution consult the user s manual Protective ground terminal Conform...

Page 11: ...erminal connect to the power line is reliable before connectes any other input and output terminal To avoid personal injury and prevent damage to the product or any product connected to this product In order to avoid possible danger the product can only be used in the specified scope Only personnel that have received professional training can execute maintenance procedures In order to prevent fire...

Page 12: ...ortation please keep the packaging and notice the relvenant transportation department and the distributor they will replace or maintain the product 2 1 2 Inspect the Accessories UTG9000T accessories power line apply for the local country region one USB four BNC cable 1 meter If the accessories are lost or damaged please contact with the distributor or the local office 2 1 3 Inspect the Instrument ...

Page 13: ... source with power line in accessories or other lines up to standard Toggle on the power switch on the back panel to operate the instrument Turn ON OFF backlight is on red when power supply in normal Press the key the backlight is on green Aferwards the screen enters function interdace after displaying start up interface In order to prevent accidentally touching ON OFF to turn off the instrument t...

Page 14: ...s etc and the channel output terminal export the signal at the same time Turn off the output function the backlight of the CH1 will be also light off the channel tab becomes grey and the channel output terminal closed Numeric Key and Utility The numeric key is used to enter numbers 0 9 decimal point symbol key and delete key Utility key is used to set multipurpose settings Direction Key The direct...

Page 15: ...hese holes Internal 10MHz output terminal Establish synchronous or external clock signal with reference frequency of 10 MHz for multiple function arbitrary waveform generators When clock source of the instrument is internal internal 10MHz output terminal outputs an internal 10MHz clock signal External 10MHz intput terminal Establish synchronization of multiple function arbitrary waveform generator...

Page 16: ...nal modulation interface This pulse string can output a designated cycle number of pulse string External Analog Modulation Output Terminal In case of AM FM PM DSB AM SUM or PWM signal if modulation is external input signal through through external analog modulation The corresponding modulation of depth frequency deviation phase deviation or duty ratio deviation is controlled by 5V signal level of ...

Page 17: ...elp symbol tap this symbol to open the help menu 3 Frequency symbol tap this symbol to open frequency counter it presents the test result Menu tab tap CH1 CH2 CH3 CH4 and Utility to make parameter and secondary function settings Highlight display Select tab will be highlight with CH color or cyan of the secondary function words with white color Output Mode continue modulate sweep burst Carrier wav...

Page 18: ...r wave AM linear or N period Wave Display Area display the current waveform it can distinguish by color or highlight of the CH tab parameter list display the current waveform parameters on the left side Note There is no waveform display area in Utility page CH Status Settings quick switch the general settings of the current channel Tap channel tab to switch output on off to enable the channel outp...

Page 19: ...quency output settings Amplitude output settings DC offset voltage settings Square wave settings Pulse wave settings DC voltage settings Ramp wave settings Noise wave settings Harmonic wave settings PRBS settings Noise superposition settings 2 3 1 Frequency Output Settings The instrument output of a sine wave is frequency 1kHz amplitude 100mVpp default setting when activating the instrument The st...

Page 20: ...ep through unit of Vpp Vrms dBm Note dBm setting only enable when Load is no HighZ mode Figure 2 5 Amplitude Settings 2 3 3 DC Offset Voltage Settings The instrument output DC offset voltage of a sine wave amplitude is 0V default setting when activating the instrument The step to set the DC offset voltage to 150mV 1 Tap Continue tab to select Sine 2 Tap Offset tab pop out the virtual numeric keybo...

Page 21: ...quare wave mode tap Amplitude tab to pop out virtual numeric keyboard to enter 1 5Vpp 2 Tap Duty tab pop out virtual numeric keyboard to enter 70 3 Tap word Duty again to step through Duty PWidth Figure 2 7 Square Settings 2 3 5 Pulse Wave Settings The duty ratio of pulse wave presents the time quantum between with threshold value of rising edge 50 decrease to the next falling edge 50 assuming tha...

Page 22: ...Settings 2 3 6 DC Voltage Settings The default value is 0V of the DC voltage The step to set DC offset voltage to 3V 1 Tap Continue tab to select DC wave mode 2 Tap Offset tab pop out the virtual numeric keyboard to enter 3V Figure 2 9 DC Settings 2 3 7 Ramp Wave Settings The symmetry presents the ramp slope is the positive of time quatunm in each cycling assuming that the waveform is not inverse ...

Page 23: ...0mVpp 1 Tap Continue tab to select Noise wave mode 2 Tap Frequency tab pop out the virtual numeric keyboard to enter 100MHz 3 Tap Amplitude tab pop out the virtual numeric keyboard to enter 300mVpp Figure 2 11 Noise Wave Settings 2 3 9 Hamonic Wave Settings UTG9000T function arbitrary waveform generator can output the designated count amplitude and phase According to the Fourier Transform theoy th...

Page 24: ...s All harmonic wave 2 times the amplitude of harmonic 4Vpp the phase of harmonic 0 1 Tap Continue tab to select Harmonic 2 Tap Frequency tab pop out the virtual numeric keyboard to enter 1MHz 3 Tap Amplitude tab pop out the virtual numeric keyboard to enter 5Vpp 4 Tap Total numbe tab pop out the virtual numeric keyboard to enter 2 5 Tap Type tab to select All 6 Tap Amplitue of harmonic wave tab po...

Page 25: ...nue tab to select Sine 2 Tap Frequency tab pop out the virtual numeric keyboard to enter 10kHz 3 Tap Amplitude tab pop out the virtual numeric keyboard to enter 2Vpp 4 Tap Noise to turn on Note Differen frequency and amplitude will effect the range of SNR The default noise superposition is 10dB Figure 2 14 Noise Superposition Settings 2 4 Secondary Function Settings Utility can set the settings of...

Page 26: ...verse OFF ON Load High resistance default 1Ω 1000kΩ Amplitude Limit OFF ON High The high limit of the output amplitude Low The low limit of the output amplitude Tap Utility Channel to make the relevant settings 1 Output Tap Output to step through ON OFF Note CH1 CH2 CH3 CH4 key can quick turn on off the chanel output function on the front panel 2 Inverse Tap INV to step through ON OFF 3 Load Tap L...

Page 27: ...wing Table 2 3 Frequency Coupling Menu Sub menu Settings Description Frequency Coupling Frequency OFF ON Type Ratio Deviation Ratio CH2 CH1 or CH4 CH3 Turn on Ratio to make the setting Deviation CH2 CH1 or CH4 CH3 Turn on Deviation to make the setting Table 2 4 Amplitude Coupling Meue Sub menu Setting Description Amplitude Coupling Amplitude OFF ON Type Ratio Deviation Ratio CH2 CH1 or CH4 CH3 Tur...

Page 28: ...rn key 2 Amplitude Coupling The amplitude coupling mode to set ratio or deviation CH1 and CH2 are each other s reference sources change one of the channel s as the reference source amplitude the other channel will adjustment automatically And always maintain the specified ratio deviation of the reference channel Ratio the specific value of CH2 CH1 Deviation the difference value of CH2 CH1 The ampl...

Page 29: ... and tap return key 4 Display Icon Turn on channel coupling the coupling icon appears on the right side of the waveform parameters display list in the home page Figure 2 15 Channel Coupling 2 4 3 Channel Merging Tap Utility Coupling to make the relevant settings In general the output terminal of CH1 2 is only export the CH1 2 waveform turn on channel merging function can output the CH1 and CH2 wav...

Page 30: ...riod ratio positive pulse and negative pulse etc And conculateing the measuring results The instrument is auto measuring the maximum minimum the average value and standard deviation Table 2 6 Frequency Counter Settings Menu Sub menu Setting Description Frequency counter Switch OFF ON Coupling AC DC To ensure accurate measurement turn on AC when the input signal frequency is high turn on DC when th...

Page 31: ...the following option IP address Mask Gateway MAC Tap Utility System to enter the Internet settings interface on the right side 1 DHCP Tap DHCP to step through OFF ON 2 IP Address IP address format nnn nnn nnn nnn the first of nnn range can set to 1 233 the second 0 255 It is suggest that you inquiry the internet administrator to get an available IP address Tap IP or push the numeric keyboard to en...

Page 32: ...y It is represented as a binary number an unsigned integer and written as a hexadecimal number 2 4 6 System Table 2 8 System Settings Menu Sub menu Setting Description System Language Chiese English Separator Comma Space None Beep OFF ON Backlight 30 40 50 60 70 80 90 100 Screen Saver OFF 5min 15 min 30 min 60 min Clock Source Internal External Clock Output OFF ON Power On Default Last Set1 Se2 Se...

Page 33: ...TL If the instrument inspect the terminal has no effect external clock screen will pop out a notice External clock is invalid And display the symbol on the top right If the instrument inspect the terminal of external clock is effect it displays the symbol 6 Clock Output Tap Clock Output to step through OFF ON Turn on the clock output the terminal of 10MHz Out for other instrument The synchronizati...

Page 34: ...el The AllChannel function can work on all the channel at the same time which the channel trigger source is turned on 10 Save Tap Save to keep the current settings as the preview settings it can select to keep into Set1 Set2 Set3 Set4 and Set5 Default Tap Default to restore the instrument to Factory settings pop out dialog box and tap it to Cancel Ok 11 About Display model name version information...

Page 35: ...n as following AM Amplitude Modulation FM Frequency Modulation PM Phase Modulation ASK Amplitude Shift Keying FSK Frequency Shift Keying 3FSK Three Frequency Shift Keying 4FSK Four Frequency Shift Keying PSK Phase Shift Keying BPSK Double Phase Shift Keying QPSK Quad Phase Shift Keying SUM Sum Modulation DSB AM Double Side Amplitude Modulation QAM Quadrature Amplitude Modulation OSK Oscillation Ke...

Page 36: ...ary wave except DC the default option is sine wave After select AM modulation tap Base to present the carrier waveform on the right side Carrier frequency settings The frequency range of the carrier wave can set differently all the default fundamental frequency is 1kHz see Table 4 1 and Table 4 2 for the frequency settings of the different carrier waveform Table 4 1 The carrier wave frequency of C...

Page 37: ...1uHz 50MHz 1uHz 40MHz Rotate multifunction knob and direction key or tap ModFreq tab to pop out visual numeric keyboard to set the carrier wave frequency Select the source of modulation UTG9000T function arbitrary waveform generator can select internal or external modulation source When turn on AM modulation the default modulation source is internal Rotate multifunction knob or tap ModSrc tab to s...

Page 38: ...oard to enter values and select unit in modulation setting interface When the modulation source is external the modulation wave and frequence will be hidden in parameter list Use the external waveform to modulating the modulation wave The intput external modulating signal frequency range is between 0Hz 50kHz Modulation depth settings Modulation depth indicates the change of the extent of amplitude...

Page 39: ...odulation mode tap CH1 Modulate AM accordingly 2 Set the parameter of modulating signal Base on the step 1 to tap ModFreq tab to pop out visual numeric keyboard to enter 200Hz 3 Set the parameter of the carrier wave signal Tap Base to select Square as the carrier waveform the default option is sine wave Tap Freq tab to pop out visual numeric keyboard to enter 10kHz Tap Ampl tab to pop out visual n...

Page 40: ... Generator 40 4 Set the modulation depth After the parameter setting of the carrier wave tap Type to return to AM modulation interface to set the depth Tap ModDepth tab to pop out visual numeric keyboard to enter 80 5 Turn on the output channel ...

Page 41: ...the AM modulation waveform viewed through an oscilloscope is shown in the figure below 3 1 2 FM Frequency Modulation In frequence modulation mode the modulated waveform is consist of the carrier wave and the modulation wave The carrier wave frequency changed by the modulation wave amplitude The modulation modes for each channel are independent it can set the same or different modulation modes to e...

Page 42: ...waveform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation Select the source of modulation UTG9000T function arbitrary waveform generator can select internal or external modulation source When turn on FM modulation the default modulation source is internal Rotate multifunction knob or tap ModSrc tab to step through internal or external ...

Page 43: ...e carrier wave The FM offset is controlled by the 5V signal level on the external analog modulation input terminal Modulation In connector on the back panel The frequency of the output at the positive signal level is greater than the carrier wave frequency at the negative signal level is less than the carrier wave frequency the lower external signal level produces less deviation For example if set...

Page 44: ...r values and select unit in modulation setting interface The frequency deviation must the carrier wave frequency if the frequency deviation is greater than fundamental frequency the function arbitrary waveform generator automatically limits the deviation to the maximum which allowed by the current fundamental frequency The sum of the frequency deviation and the carrier wave frequency must the maxi...

Page 45: ...oard to enter 2kHz 3 Set the parameter of the carrier wave signal Tap Base to select sine as the carrier wave the default option is sine so this step has no change Tap Freq to pop out visual keyboard to enter 10kHz 4 Set the frequency deviation After the parameter setting of the carrier wave tap Type to return to modulation interface to set the frequency deviation ...

Page 46: ...n the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 key and CH1 tab backlight light on indicating that the channel 1 output is turned on The shape of the FM modulation waveform viewed through an oscilloscope is shown in the figure below ...

Page 47: ...can set the same or different modulation modes to each channel Select PM modulation Tap CH1 Modulate PM to turn on the AM modualtion the instrument output the moduluated waveform according to the current setting of the modulation wave and the carrier wave Select the carrier waveform The carrier wave of PM sine square ramp pulse and arbitrary wave except DC the default option is sine wave After sel...

Page 48: ...ernal or external modulation source in PM modulation setting interface 1 Internal source When the modulation source is internal the modulation waveform can be sine square rising ramp falling ramp arbitrary and noise wave The default wave is sine wave Whturn on FM modulation ModWave present Sine rotate multifunction knob or tap word Sine to select the waveform in modulation setting interface Square...

Page 49: ...es and select unit in PM modulation setting interface When the modulation source is external the modulation wave and frequence will be hidden in parameter list Use the external waveform to modulating the carrier wave The intput external modulating signal frequency range is between 0Hz 50kHz Modulated phase deviation settings Phase deviation indicates the phase deviation changes between in the PM m...

Page 50: ... signal Base on the step 1 to tap ModFreq tab to pop out visual numeric keyboard to enter 200Hz 3 Set the parameter of the carrier wave signal Tap Base to select sine as the carrier wave the default option is sine so this step has no change Tap Freq to pop out visual keyboard to enter 900Hz ...

Page 51: ...o pop out visual numeric keyboard to enter 200 5 Turn on the output channel Tap channel Out botton to ON or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 key and CH1 tab backlight light on indicating that the channel 1 ...

Page 52: ... changing the amplitude of carrier signal And according to the logical of modulation signal to output the carrier signals with different amplitude The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Select ASK modulation Tap CH1 Modulate ASK to turn on ASK modulation the instrument output the moduluated waveform according to the c...

Page 53: ...waveform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation Select the source of modulation UTG9000T function arbitrary waveform generator can select internal or external modulation source When turn on FM modulation the default modulation source is internal Rotate multifunction knob or tap ModSrc tab to step through internal or external ...

Page 54: ...mplitude is less than the current carrier amplitude when the external input is high Modulation waveform frequency settings When the modulation source is internal it can set the modulation waveform frequency and the frequency range is between 1uHz 2MHz the default range is 100Hz Rotate multifunction knob push direction key or tap ModFreq tab to pop out visual numeric keyboard to enter values and se...

Page 55: ...s the carrier wave the default option is sine so this step has no change Tap Freq to pop out visual keyboard to enter 15kHz Tap Ampl to pop out visual keyboard to enter 2Vpp 3 Set the parameter of the modulating signal After the parameter setting of the carrier wave tap Type to return to modulation interface to set the frequency ...

Page 56: ...n the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 key and CH1 tab backlight light on indicating that the channel 1 output is turned on The shape of the ASK modulation waveform viewed through an oscilloscope is shown in the figure below ...

Page 57: ...quency or hopping frequency The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Select FSK modulation Tap CH1 Modulate FSK to turn on FSK modulation the instrument output the moduluation waveform according to the current setting Select the carrier waveform The carrier wave of FSK sine square ramp pulse and arbitrarty wave except D...

Page 58: ...K modulation setting interface 1 Internal source When the modulation source is internal the modulation wave is a square sine with duty ratio 50 buit in and can t adjustable It can select shift frequency between the carrier wave frequency and hopping frequency by the frequency settings 2 External source When the modulation source is external rate will be hidden in paramteter list Use the external w...

Page 59: ... the carrier wave frequency and hopping frequency Turn on FSK modulation mode to set the frequency the range is between 1uHz 2MHz default 100Hz Rotate multifunction knob push direction key or tap ModFreq tab to pop out visual numeric keyboard to enter values and select unit in modulation setting interface Comperhensive example In FSK modulation mode set an internal 2kHz 1Vpp sine wave as a carrier...

Page 60: ...et the modulating frequency Tap ModFreq to pop out visual keyboard to enter 200Hz Tap HopFreq1 to pop out visual keyboard to enter 800Hz 4 Turn on the output channel Tap channel Out botton to ON or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to...

Page 61: ...orm generator can move between among three preset frequencies carrier frequency and two hopping frequency According to the logical level of the modulating signal to output the carrier frequency or hopping frequency The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Select 3FSK modulation Tap CH1 Modulate 3FSK to turn on 3FSK modu...

Page 62: ...nt the carrier waveform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation Hopping frequency settings Rotate multifunction knob push direction key or tap HopFreq1 and HopFreq2 tab to pop out visual numeric keyboard to enter values and select unit in 3FSK modulation setting interface The range of hopping frequency depends on carrier wavef...

Page 63: ...al numeric keyboard to enter values and select unit in modulation setting interface Comperhensive example In FSK modulation mode set an internal 2kHz 1Vpp sine wave as a carrier wave set HopFreq1 to 1kHz set HopFreq2 to 5kHz make the carrier wave and hopping frequence shift between within 100Hz frequency The settings steps as following 1 Turn on FSK modulation mode tap CH1 Modulate 3FSK accordingl...

Page 64: ...ncy Tap HopFreq1 to pop out visual keyboard to enter 1kHz Tap HopFreq2 to pop out visual keyboard to enter 5kHz 4 Turn on the output channel Tap channel Out botton to ON or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 ...

Page 65: ...rm generator can move between among three preset frequencies carrier frequency and three hopping frequency According to the logical level of the modulating signal to output the carrier frequency or hopping frequency The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Select 4FSK modulation Tap CH1 Modulat 4FSK to turn on 4FSK modu...

Page 66: ...e carrier waveform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation Hopping frequency settings Rotate multifunction knob push direction key or tap HopFreq1 HopFreq2 and HopFreq3tab to pop out visual numeric keyboard to enter values and select unit in 4FSK modulation setting interface The range of hopping frequency depends on carrier wa...

Page 67: ...nd select unit in modulation setting interface Comperhensive example In 4FSK modulation mode set an internal 500Hz 1Vpp sine wave as a carrier wave set HopFreq1 to 2kHz set HopFreq2 to 5kHz set HopFreq2 to 10kHz make the carrier wave and hopping frequence shift between within 100Hz frequency The settings steps as following 1 Turn on FSK modulation mode tap CH1 Modulate 4FSK accordingly 2 Set the p...

Page 68: ...2kHz Tap HopFreq2 to pop out visual keyboard to enter 5kHz Tap HopFreq3 to pop out visual keyboard to enter 10kHz 4 Turn on the output channel Tap channel Out botton to ON or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH...

Page 69: ... carrier phase and hopping phase According to the logical level of the modulating signal to output the carrier frequency or hopping frequency The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Select PSK modulation Tap CH1 Modulate PSK to turn on PSK modulation the instrument output the moduluated waveform according to the curren...

Page 70: ...eform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation Select the source of modulation UTG9000T function arbitrary waveform generator can select internal or external modulation source When turn on PSK modulation the default modulation source is internal Rotate multifunction knob or tap ModSrc tab to step through internal or external mo...

Page 71: ...frequency settings When the modulation source is internal it can select shift frequency between the carrier wave frequency and hopping frequency Turn on PSK modulation mode to set the frequency the range is between 1uHz 2MHz default 100Hz Rotate multifunction knob push direction key or tap ModFreq tab to pop out visual numeric keyboard to enter values and select unit in modulation setting interfac...

Page 72: ... signal Tap Base to select sine as the carrier wave the default option is sine so this step has no change Tap Freq to pop out visual keyboard to enter 2kHz Tap Ampl to pop out visual keyboard to enter 2Vpp 3 Set the modulating frequency and phase After the parameter setting of the carrier wave tap Type to return to modulation interface to set the modulating frequency and phase ...

Page 73: ...N or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 key and CH1 tab backlight light on indicating that the channel 1 output is turned on The shape of the PSK modulation waveform viewed through an oscilloscope is shown in...

Page 74: ...annel are independent it can set the same or different modulation modes to each channel Select BPSK modulation Tap CH1 Modulate BPSK to turn on BPSK modulation the instrument output the moduluated waveform according to the current setting of the modulation wave and the carrier wave Select the carrier waveform The carrier wave of BPSK sine square ramp pulse and arbitrarty wave except DC the default...

Page 75: ...on mode the default PN code is PN3 Rotate multifunction knob or tap PN Code PN3 to select different PN code BPSK rate settings In BPSK modulation mode it can set the shift frequency between in the carrier phase and the modulating phase The BPSK bitrate range cam set to 1μbps 2Mbps the default range is 100bps Rotate multifunction knob push direction key or tap Bitrate tab to pop out visual numeric ...

Page 76: ... BPSK modulating phase range can set to 0º 360º the settings steps is the same as Phase1 Comperhensive example In BPSK modulation mode set an internal 2kHz 2Vpp sine wave as a carrier wave set the initial phase to 90 1kbps shift frequence between in the carrier wave phase and the modulating wave phase set PN Code to PN15 The settings steps as following 1 Turn on BPSK modulation mode tap CH1 Modula...

Page 77: ...sual numeric keyboard to enter 1kbps Phase use the default value phase1 0 phase2 90 Tape PN Code PN3 to select PN15 4 Turn on the output channel Tap channel Out botton to ON or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF ...

Page 78: ...in four preset phase carrier phase and three modulating phase According to the logical level of the modulating signal to output the carrier phase or modulating phase The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Select QPSK modulation Tap CH1 Modulate QPSK to turn on FSK modulation the instrument output the moduluated wavefo...

Page 79: ...ct QPSK modulation tap Base to present the carrier waveform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation The PN code settings UTG9000T function arbitrary waveform generator can select the internal source Turn on QPSK modulation mode the default PN code is PN3 Rotate multifunction knob or tap PN Code PN3 to select different PN code ...

Page 80: ...4 is the modulating phase it means the phase changes between the QPSK modulated wave and the carrier wave phase The QPSK modulating phase range can set to 0º 360º Rotate multifunction knob push direction key or tap Phase1 Phase2 Phase3 tab to pop out visual numeric keyboard to enter values and select unit in modulation setting interface Comperhensive example In QPSK modulation mode set an internal...

Page 81: ...carrier wave the default option is sine so this step has no change Tap Freq to pop out visual keyboard to enter 2kHz Tap Ampl to pop out visual keyboard to enter 2Vpp 3 Set QPSK bitrate and the modulating phase After the parameter setting of the carrier wave tap Type to return to modulation interface to set the QPSK bitrate and phase ...

Page 82: ...l Tap channel Out botton to ON or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 key and CH1 tab backlight light on indicating that the channel 1 output is turned on The shape of the BPSK modulation waveform viewed throu...

Page 83: ...tarts oscillation output is stopped when internal crystal oscillator stops oscillation The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Select OSK modulation Tap CH1 Modulate QSK to turn on OSK modulation the instrument output the moduluated waveform according to the current setting of the modulation wave and the carrier wave S...

Page 84: ...gh internal or external modulation source in OSK modulation setting interface 1 Internal trigger source When the modulation source is internal the modulation wave is sine wave Use OSK rate to control the phase relationship of oscillation starting and stopped 2 External trigger source When the modulation source is external rate will be hidden in paramteter list Use the external waveform to modulati...

Page 85: ...ulated wave phase and the carrier wave phase in internal trigger source The rate range can set to 1uHz 2MHz the default rate is 100Hz Rotate multifunction knob push direction key or tap ModRate tab to pop out visual numeric keyboard to enter values and select unit in modulation setting interface Comperhensive Example In OSK modulation mode set an internal 2kHz 2Vpp sine wave as a carrier wave set ...

Page 86: ...mpl to pop out visual keyboard to enter 2Vpp 3 Set the modulating rate After the parameter setting of the carrier wave tap Type to return to modulation interface to set the modulating rate Tap ModRate to pop out visual keyboard to enter 100Hz the default 100Hz Tap OscTime to pop out visual keyboard to enter 1us ...

Page 87: ...k the channel tab at the bottom of the screen to step through ON OFF CH1 key and CH1 tab backlight light on indicating that the channel 1 output is turned on The shape of the PSK modulation waveform viewed through an oscilloscope is shown in the figure below 3 1 12 SUM Sum Modulation Sum modulation is consist of the carrier wave and the modulated wave The output waveform is sum of the amplitude of...

Page 88: ... on the AM modualtion the instrument output the moduluated waveform according to the current setting of the modulation wave and the carrier wave Select the carrier waveform The carrier wave of SUM sine square ramp pulse harmonic noise and arbitrary wave except DC the default option is sine wave After select SUM modulation tap Base to present the carrier waveform on the right side Carrier wave freq...

Page 89: ...to step through internal or external modulation source in AM modulation setting interface Square wave duty ratio 50 Rising ramp wave symmetry 100 Falling ramp wave symmetry 0 Arbitrary wave the length of arbitrary wave limit at 2kpts by the method of select point automatically Noise wave white gaussian noise 2 External source When the modulation source is external the modulation wave and frequence...

Page 90: ... to express SUM modulation depth range can set between 0 100 the default range is 100 If the modulation depth is 0 then output the carrier wave if the modulation depth is 100 output the modulated wave Rotate multifunction knob and direction key or tap ModDepth tab to pop out visual numeric keyboard to enter values When the modulation source is external the output amplitude is controlled by the 5V ...

Page 91: ... enter 200mVpp Tap Duty tab to pop out visual numeric keyboard to enter 45 3 Set the modulation frequency and depth After the parameter setting of the carrier wave tap Type to return to SUM modulation interface to set the frequency and depth Tap ModFreq tab to pop out visual numeric keyboard to enter 1kHz Tap ModDepth tab to pop out visual numeric keyboard to enter 80 ...

Page 92: ...slo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 and CH1 tab backlight light on indicating that the channel 1 output is turned on The shape of the AM modulation waveform viewed through an oscilloscope is shown in the figure below 3 1 13 DSB AM Double Side Band Amplitude Modulation ...

Page 93: ...DSB AM sine square ramp pulse and arbitrary wave except DC the default option is sine wave After select DSB AM modulation tap Base to present the carrier waveform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation Select the source of modulation UTG9000T function arbitrary waveform generator can select internal or external modulation sou...

Page 94: ...ulation wave and frequence will be hidden in parameter list Use the external waveform to modulating the carrier wave The DSB AM depth is controlled by the 5V signal level on the external analog modulation input terminal Modulation In connector on the back panel For example if set the modulation depth to 100 the output of DSB AM amplitude is maximum when the external modulation signal is 5V the out...

Page 95: ...ine wave signal with frequency 2kHz amplitude2Vpp as the carrier wave signal the settings steps as following 1 Turn on AM modulation mode tap CH1 Modulate AM accordingly 1 Set the parameter of the carrier wave signal Tap Base to select Square as the carrier waveform the default option is sine wave Tap Freq tab to pop out visual numeric keyboard to enter 2kHz Tap Ampl tab to pop out visual numeric ...

Page 96: ...visual numeric keyboard to enter 1kHz 4 Turn on the output channel Tap channel Out to ON or push CH1 on the front panel to quick set the output channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 and CH1 tab backlight light on indicating that the channel 1 output is turned on The ...

Page 97: ...28 and QAM256M The modulation modes for each channel are independent it can set the same or different modulation modes to each channel Note It is recommened to use the 10MHz reference output signal of this instrument as the input reference clock of the demodulation instrument or input the demodulation instrument reference clock as the signal clock Accurate the signal demodulation and eliminate the...

Page 98: ... present the carrier waveform on the right side Carrier wave frequency settings Refer to The carrier wave frequency settings of AM modulation Modulation type The modulation type which is the division of the constellation changed by the modulation types Rotate multifunction knob or tap IQMap QAM4 to step through QAM4 QAM8 QAM16 QAM32 QAM64 QAM128 and QAM256M ...

Page 99: ...am set to 1μbps 2Mbps the default range is 100bps Rotate multifunction knob push direction key or tap Bitrate tab to pop out visual numeric keyboard to enter values and select unit in modulation setting interface Comperhensive example In QAM modulation mode set an internal 2kHz 2Vpp sine wave as a carrier wave set the rate to 100bps modulation type QAM64 PN code PN7 The settings steps as following...

Page 100: ...op out visual keyboard to enter 2Vpp 3 Set the modulation parameter After the parameter setting of the carrier wave tap Type to return to modulation interface to set parameter Tape IQMap QAM4 to select QAM64 Tape PN Code PN3 to select PN7 Tape Bitrate to pop out visual numeric keyboard to enter 100bps the default 100bps ...

Page 101: ...ce tap Utility Channel Output or double click the channel tab at the bottom of the screen to step through ON OFF CH1 key and CH1 tab backlight light on indicating that the channel 1 output is turned on The shape of the QAM modulation waveform viewed through an oscilloscope is shown in the figure below 3 1 15 PWM Pulse Width Modulation In pulse width modulation mode the modulated waveform is consis...

Page 102: ...hannel Select PWM modulation Tap CH1 Modulate PWM to turn on PWM modulation the instrument output the moduluated waveform according to the current setting of the modulation wave and the carrier wave Select the carrier waveform PWM carrier waveform can only be pulse wave After select PWM modulation tap Base to present the carrier waveform on the right side Carrier wave frequency settings Refer to T...

Page 103: ...tting interface Square wave duty ratio 50 Rising ramp wave symmetry 100 Falling ramp wave symmetry 0 Arbitrary wave the length of arbitrary wave limit at 4kpts by the method of select point automatically Noise wave white gaussian noise 2 External source When the modulation source is external the modulation wave and frequence will be hidden in parameter list Use the external waveform to modulating ...

Page 104: ...lt is 49 999825 Rotate multifunction knob push direction key or tap DutyDev tab to pop out visual numeric keyboard to enter values and select unit in FM modulation setting interface Duty ratio deviation is deviation of modulated waveform from the current carrier duty ratio express by Duty ratio deviation can t exceed the duty ratio of the current pulse wave The sum of duty ratio deviation and the ...

Page 105: ...enter 10kHz Tap Ampl to pop out visual numeric keyboard to enter 2Vp Tap REdge to pop out visual numeric keyboard to enter 100ns Tap FEdge to pop out visual numeric keyboard to enter 100ns 3 Set the modulating parameter After the parameter setting of the carrier wave tap Type to return to modulation interface to set the parameter Tap ModFreq to pop out visual numeric keyboard to enter 1kHz Tap Dut...

Page 106: ...utput channel and it can aslo turn on at the Utility interface tap Utility Channel Output or double click the channel tab bar at the bottom of the screen to step through ON OFF CH1 key and CH1 tab bar backlight light on indicating that the channel 1 output is turned on The shape of the FM modulation waveform viewed through an oscilloscope is shown in the figure below ...

Page 107: ...lation modes for each channel are independent it can set the same or different modulation modes to each channel 3 2 1 Select Frequency Sweep 1 Turn on frequency sweep mode tap CH1 Sweep accordingly the instrument output the current frequency sweep waveform 2 Select the frequency sweep waveform Tape Base tap Base to present the carrier waveform on the right side See the interface figure as below 3 ...

Page 108: ...of frequency sweep mode is low from starting point to midpoint of frequency sweep time and high from midpoint to end of frequency sweep time By default starting frequency is 1kHz and stop frequency is 1Mz but the range of starting and stop frequency can vary with frequency sweep waveform See the Table 4 1 4 2 Carrier frequency settings of AM 3 2 3 Frequency Sweep Mode UTG9000T has four frequency s...

Page 109: ...ncy sweep can be internal external external rising falling edge or manual Rotate multifunction knob or tap ModSrc to step through Internal External ExREdge ExFEdge and Maunal 1 The waveform generator will output a succession frequency sweep when in internal trigger source The rate controlled by frequency sweep time 2 The waveform generator will trigger by the external digital modulation terminal F...

Page 110: ...L level The default trigger output is OFF Rotate multifunction knob or tap TrigOut to step through Close REdge and DEdge Internal trigger source the waveform generator outputs a square wave with duty ratio of 50 from external digital modulation terminal FSK Trig connector when frequency sweep starts Trigger period depends on designated frequency sweep time Manual trigger source the waveform genera...

Page 111: ...ge represents the rising edge output trigger signal falling edge represents the falling edge output trigger singal and the default is rising edge 3 2 8 Comperhensive example In frequency sweep mode set a square wave with amplitude 1Vpp duty ratio 50 as frequency sweep the frequency sweep type is linear set the starting frequency to 1kHz stop frequency 50kHz and frequency sweep time to 2ms use the ...

Page 112: ... Linear as the way of frequency sweep default Tap Start tab to pop out visual numeric keyboard to enter 1kHz Tap Stop tab to pop out visual numeric keyboard to enter 50kHz Tap DwellTime tab to pop out visual numeric keyboard to enter 2ms 3 Turn on the output channel Tap channel Out botton to ON or push CH1 key on the front panel to quick set the output channel and it can aslo turn on at the Utilit...

Page 113: ...e string It has three trigger output mode internal external and manual And offer N cycle gating and infinite pulse string It can generate pulse train for sine wave square wave ramp wave pulse wave arbitrary wave except DC and noise only applicable to gating pulse train The modulation modes for each channel are independent it can set the same or different modulation modes to each channel 3 3 1 Sele...

Page 114: ...ting modes waveform frequency defines the signal frequency during period of pulse train In N cycle mode output the pulse train at a specified cycle times and waveform frequency In gating mode use waveform frequency to output the pulse string when the trigger source is high level Note Waveform frequency is different from pulse string period Pulse string is used to specify the interval between pulse...

Page 115: ...lso output by the external digital modulation terminal FSK Trig connector This terminal can not be the external trigger input and internal trigger output at the same time 2 Gating mode Tap Gate to turn on gating mode trigger source trigger output trigger period and cycle number will be hidden in parameter list It can only use the external trigger source the waveform generator trigged by the extern...

Page 116: ...l trigger out option will be hidden in parameter list Because the trigger output is also output by the external digital modulation terminal FSK Trig connector This terminal can not be the external trigger input and internal trigger output at the same time 3 3 3 Initial Phase of Burst Start phase of burst cab set to 0 360 the default start phase is 0 Rotate multifunction knob or tap Phase tab to po...

Page 117: ...crease the designated period to output a designated number of cycles 3 3 5 Counting of Burst In N cycle mode pulse string counting is used to specify the number of waveform period It is in the range of 1 50000 periods and 2 by default In N cycle mode rotate multifunction knob push direction key or tap Cycle tab to pop out visual numeric keyboard to enter values and select unit Cycle number Trigger...

Page 118: ...OFF Rotate multifunction knob or tap TrigOut to step through Close REdge and DEdge Internal trigger source the waveform generator outputs a square wave with duty ratio of 50 from external digital modulation terminal FSK Trig connector when burst mode starts The waveform frequency is equal to pulse string period Manual trigger source the waveform generator outputs a pulse wide than 1μs from externa...

Page 119: ...p as pulse string wave burst mode set to N cycle pulse string period 15ms cycle number 2 the settings steps as following 1 Turn on burst mode tap CH1 Burst NCycle accordingly 2 Select burst waveform Tap Base o select sine wave as the carrier waveform the default option is sine so this step has no change Note If the setting interface display Freq this means it can only change the frequency paramete...

Page 120: ... Period to pop out visual numeric keyboard to enter 5ms Tap Ampl to pop out visual numeric keyboard to enter500mVpp 3 Set burst period and cycle After select burst waveform tap Type to turn back to burst interface Tap parameter bar to enter value and select unit ...

Page 121: ...the bottom of the screen to step through ON OFF CH1 key and CH1 tab bar backlight light on indicating that the channel 1 output is turned on The shape of the burst waveform viewed through an oscilloscope is shown in the figure below 3 4 Output Abitrary Waveform UTG9000T hs store more than 200 standard waveform types in non volatile memory See Table 4 4 Built in Arbitrary Waveform The instrument ca...

Page 122: ...536k points and sampling rate The waveform generator outputs waveform points one by one with this frequency It can prevent loss of important waveform point In DDS default mode the waveform generator output arbitrary waveform by automatic interpolation or select point as fixed length 8 192 points and the frequency of parameter list Rotate multifunction knob or tap Mode tab to step through DDS and P...

Page 123: ...l Common AbsSine bsv Tap Load to turn off window automatically arbitrary waveform selection is finished Note UTG9000T supports the length less than 64 points with file format csv or bsv Table 4 4 Buit in Arbitrary Waveform Type Name Description Common 15 types Sin Sine function Square Sqaure waveform Ramp Sawtooth waveform ...

Page 124: ...ting Time vibration velocity curve Butterworth Butterworth filter Chebyshev1 Type I Chebyshev filter Chebyshev2 Type II Chebyshev filter Combin Compound function CPulse C Pulse signal CWPulse CW pulse signal DampedOsc Damped vibrarion Time displacement curve DualTone Dual tone signal Gamma Gamma signal GateVibar Gate self excited osclliation signal LFMPulse Linear frequency modulation pulse signal...

Page 125: ... error function ErfInv Inverse error function ExpFall Exponential falling function ExpRise Exponential rising function Gammaln Natural logarithm of Gamma function Gauss Gaussian distribution Normal distribution HaverSine Haversed sine Laguerre Quartic Laguerre polynomial Laplace The Laplace distribution Legend Quintic Legendre Polynomials Log Denary logarithm function LogNormal Logarithmic normal ...

Page 126: ...Tens3 Transcutaneous electric nerve stimulation waveform 3 Standard 17 types Ignition Ignition waveform of automobile internal combustion engine ISO16750 2 SP Profile map of automobile starting oscillation ISO16750 2 Starting1 Automobile starting voltage waveform 1 ISO16750 2 Starting2 Automobile starting voltage waveform 2 ISO16750 2 Starting3 Automobile starting voltage waveform 3 ISO16750 2 Sta...

Page 127: ...e distribution Surge Surge signal Trigonome 21 types CosH Hyperbolic cosine CosInt Cosine integral Cot Cotangent function CotHCon Concave hyperbolic cotangent CotHPro Convex hyperbolic cotangent CscCon Concave cosine CscPro Convex cosine CotH Hyperbolic cotangent CscHCon Concave hyperbolic cosecant CscHPro Convex hyperbolic cosecant RecipCon Reciprocal of the depression RecipPro Reciprocal of the ...

Page 128: ... arc secant function ASecPro Convex arc secant function ASecH Arc hyperbolic secant function ASin Arcsin function ASinH Arc hyperbolic sine function ATan Arctan function ATanH Arc hyperbolic tangent function Noise 6 types NoiseBlue Blue noise NoiseBrown Brown noise red noise NoiseGray Gray noise NoisePink Pink noise NoisePurple Purple noise Noisewhite White noise Window 17 types Bartlett Bartlett ...

Page 129: ...ex Frequency B spline function Complex Gaussian Complex Gaussian function Complex Morlet Complex Morlet wavelet Complex Shannon Complex Shannon function Mexican hat Mexican hat wavelet Meyer Meyer wavelet Morlet Morlet wavelet Other 3 types ABA_1_1 ABA_1_2 ALT_03 ALT_04 ALT_05 AUDIO circle Circle modulaion COIL_2_1 COIL_2_2 DC_04 diamond Diamond modulaion ECT_1_2 EGR_2 EGR_3_2 EST_03_2 Heart Heart...

Page 130: ... Programming Guide The created arbitrary waveform can import to the function arbitrary waveform generator 3 5 Output Digital Protocol The waveform generator can output three types of protocol conding IIC SPI and UART TTL The corresponding protocol parameters can be set in different protocol modes Turn on output terminal on the front panel to export the corresponding signal 3 5 1 SPI Protocol The w...

Page 131: ...d to enter values and select unit The clock range can set to 1Hz 50MHz Data format setting The data format of SPI can set by user s need It have two format types hexadecimal and character In SPI mode tap Format to step through HEX and Char Set data Tap blank input box on the right side to pop out visual numeric keyboard to enter bits and tap return key to complete an entry ...

Page 132: ...ndType to select Manual mode the default is Auto tap Send on the right side to send the set waveform Set interval time If the send mode is auto set the interval time of data send according to the actual condition Tap Interval to pop out visual numeric keyboard to enter values and select unit The interval time range can set to 20ns 1000s Comprehensive example In SPI mode set hexadecimal 13 21 34 55...

Page 133: ... OFF to turn on the output function CH2 is SPI SCLK CH3 is SPI CS CH4 is SPI MOSI See the figure as below 3 5 2 IIC Protocol The waveform generator can create configurable protocol signal parameters in IIC protocol mode Select IIC mode Tap Utility Digital IIC accordingly turn on IIC mode to output the current IIC protocol signal ...

Page 134: ...d to enter values and select unit The clock range can set to 1Hz 50MHz Data format setting The data format of IIC can set by user s need It have two format types hexadecimal and character In IIC mode tap Format to step through HEX and Char Set data Tap blank input box on the right side to pop out visual numeric keyboard to enter bits and tap return key to complete an entry ...

Page 135: ... Comprehensive example In IIC mode set the output address to 10 bit values is 65 clock 500kHz select decimalism 17 19 29 29 31 as the data format interval time 5ms the setting steps as following 1 Turn on IIC mode Tap Utility Digital IIC accordingly 2 Set data parameter Tap Clock to pop out visual numeric keyboard to enter 15kHz Tap Interval to pop out visual numeric keyboard to enter 5ms Tap blan...

Page 136: ...rotocol signal Set baud rate Baud rate can set by user s need In UART mode tap Baudrate to pop out visual numeric keyboard to enter values and select unit The baud rate range can set to 1 1000000 Data format setting The data format of UART can set by user s need It have two format types hexadecimal and character In UART mode tap Format to step through HEX and Char Set data Tap blank input box on t...

Page 137: ...p Baudrate to pop out visual numeric keyboard to enter values Set bit It can set different bit in UART mode data have 4 5 6 7 8 bits The default bit is 8 Tap Baudrate to pop out visual numeric keyboard to enter values Set stop bit It can set different stop bit in UART mode Tap Stop to step through 1bit and 2bit The default is 1bit Set verify mode Tap Verify to step through None Even Odd ...

Page 138: ...ms the settings step as following 1 Turn on IIC mode Tap Utility Digital UART accordingly 2 Set data parameter Tap Baudrate to pop out visual numeric keyboard to enter 4800 Tap Data to pop out visual numeric keyboard to select 8bit Tap input box to pop out visual numeric keyboard to enter 2ms Data format and stop bit use the default option 3 Turn on output function Tap OFF to turn on the output fu...

Page 139: ...UTG9000T Series Function Abitrary Waveform Generator 139 ...

Page 140: ...UTG9000T Series Function Abitrary Waveform Generator 140 ...

Page 141: ... 5 If the instrument still can t work please contact with the dealer or local office for product maintenance service 4 2 No Waveform Output In correct setting but the instrument has no waveform output display 1 Inspect whether BNC cable and the output terminal is connected well 2 Inspect button whether CH1 CH2 CH3orCH4is turned on 3 Keep the current settings into USB and then push Factory Setting ...

Page 142: ...UTG9000T Series Function Abitrary Waveform Generator 142 ...

Page 143: ...of warranty To arrange repair or ask for the whole warranty please contact with the nearest sales or maintenance department of UNI T Except warranties provided in the outline or other applicable warranties UNI T does not provide any other express or implied warranties including but not limited to any implied warranties about tradability and applicability to special purpose of the product In any ca...

Page 144: ...UTG9000T Series Function Abitrary Waveform Generator 144 ...

Page 145: ...t Off Channel output inverse Off Amplitude limit Off Upper amplitude limit 1V Lower amplitude limit 1V Fundamental Wave Frequency 1kHz Amplitude 100mVpp DC offset 0mV Initial phase 0 Duty ratio of square wave 50 Degree of symmetry of ramp wave 100 Duty ratio of pulse wave 50 Rising edge of pulse wave 1 4ns Falling edge of pulse wave 1 4ns Arbitrary Wave Built in arbitrary wave ACos Play mode DDS A...

Page 146: ...on Modulation source Internal Modulation wave Sine wave Modulation frequency 100Hz Phase deviation 180 PWM Modulation Modulation source Internal Modulation wave Sine wave Modulation frequency 100Hz Duty ratio deviation 49 999825 ASK Modulation Modulation source Internal ASK rate 100Hz FSK Modulation Modulation source Internal FSK rate 100Hz Hopping frequency 100kHz PSK Modulation Modulation source...

Page 147: ...ource PN3 QPSK rate 100Hz Phase 1 0 Phase 2 90 Phase 3 180 Phase 4 270 OSK Modulation Modulation source Internal Oscillation time 2ms OSK rate 100Hz DSB AM Modulation Modulation source Internal Modulation wave Sine wave Modulation frequency 100Hz QAM Modulation Constellation QAM4 Coding code PN3 QAM rate 100bps SUM Modulation Modulation source Internal Modulation wave Sine wave Modulation frequenc...

Page 148: ...ternal Trigger output OFF Trigger edge Rising edge Pulse string Mode of pulse string N cycle Initial phase 0 Burst period period of pulse string 5ms Cycle number 2 Gated polarity Positive Trigger source Internal Trigger output OFF Trigger edge Rising edge System Parameter IP type DHCP Clock source Internal Clock output OFF Sound of buzzer ON Separator of numbers Backlight 100 Language Depend on fa...

Page 149: ...z 200MHz 120MHz Sampling rate 2 5GSa s CH1 and CH2 625MSa s CH3 and CH4 Waveform Sine Square Ramp Pulse Harmonic Noise PRBS DC Arbitrary wave Mode Continue Modulate Sweep Burst Frequency counter Protocol Modulation type AM FM PM DSBAM QAM ASK FSK 3FSK 4FSK PSK BPSK QPSK OSK PWM SUM Waveform Characteristic Sine Wave Frequency range 1μHz 600MHz 1μHz 200MHz 1uHz 400MHz 1μHz 160MHz 1uHz 200MHz 1μHz 12...

Page 150: ... 3ns 5ns 6ns 7ns Overshoot typical value 2 1MHz 1 Vpp 50Ω load Duty ratio 0 000001 99 999999 Symmetry duty ratio of 50 1 of period 4ns Shake typical value 100 ps 1Vpp 50Ω load Ramp Wave Frequency range 1μHz 30MHz 1μHz 10MHz 1uHz 20MHz 1μHz 8MHz 1uHz 10MHz 1μHz 3MHz Resolution 1uHz Non linearity 1 of peak output typical value 1kHz 1Vpp symmetry 100 Symmetry 0 0 100 0 Pulse Wave Frequency range 1μHz...

Page 151: ...ts CH3 and CH4 Vertical resolution 16bits including symbol Shake 150ps Nonvolatile storage more than 200 waveform PRBS Wave Bitrate 1μbps 120Mbps 1μbps 60Mbps 1μbps 80Mbps 1μbps 40Mbps 1μbps 40Mbps 1μbps 20Mbps Edge time 2 6ns 1000s CH1 and CH2 4 2ns 1000s CH3 and CH4 PN code PN3 PN5 PN7 PN9 PN11 PN13 PN15 PN17 PN21 PN23 PN25 PN27 PN29 PN31 PN33 Overlay amplitude of noise noise voltage 1Vrms Harmo...

Page 152: ...o 1kHz sine wave 1Vpp 50Ω 10MHz 0 1dB 160MHz 0 2dB 350MHz 0 4dB 500MHz 0 8dB 600MHz 0 8dB Output Waveform Impedance 50Ω typical vaule Protection Channel protection Modualtion Type AM Modualtion Carrier wave Sine square pulse ramp arbitrary wave Source Internal external Modulation wave Sine square rising ramp falling ramp noise arbitrary wave Modulation frequency 1μHz 2MHz 1μHz 1MHz Modulation dept...

Page 153: ...frequency 1μHz 2MHz 1μHz 1MHz Hopping frequency 1 1μHz 600MHz 1μHz 200MHz 1uHz 400MHz 1μHz 160MHz 1uHz 200MHz 1μHz 120MHz 3FSK Modualtion Carrier wave Sine square pulse ramp arbitrary wave Source Internal external Modulation wave 50 duty ratio of square wave Modulation frequency 1μHz 2MHz 1μHz 1MHz Hopping frequency 1 1μHz 600MHz 1μHz 200MHz 1uHz 400MHz 1μHz 160MHz 1uHz 200MHz 1μHz 120MHz Hopping ...

Page 154: ...0 duty ratio of square wave Bitrate 1μbps 2Mbps 1μbps 1Mbps Hopping phase 0 00 360 00 QPSK Modualtion Carrier wave Sine square ramp arbitrary wave PN code PN3 PN5 PN7 PN9 PN11 PN13 PN15 PN17 PN21 PN23 PN25 PN27 PN29 PN31 PN33 Modulation wave 50 duty ratio of square wave Bitrate 1μbps 2Mbps 1μbps 1Mbps Hopping phase 1 0 00 360 00 Hopping phase 2 0 00 360 00 Hopping phase 3 0 00 360 00 OSK Modualtio...

Page 155: ...rier wave Sine square ramp pulse arbitrary wave Type Linear logarithm stepping list sweep Frequency sweep time 1ms 500s Trigger source Internal external rising edge external falling edge manual Trigger output Close rising edge falling edge Step 2 2048 steps Listfile Maximum 2048 frequency points for a single file Frequency range is accordance with fundamental wave range Burst Carrier wave Sine squ...

Page 156: ...bytes UART Interface CH4 TX Amplitude 1mV 10V Baud rate 1 1000000 customized Data bit 4 5 6 7 8 Stop bit 1bit 2bit Verify bit None even odd Send way Auto manual Interval time 20ns 1000s in auto mode of send way Data format Hexadecimal character Data length Maximum 2048 bytes Channel Coupling Merge Frequency coupling Ratio 0 00001 10000 deviation 600MHz 600MHz 200MHz 200MHz 400MHz 400MHz 160MH 160M...

Page 157: ...L Input impedance 10kΩ typical value DC coupling Lock time 1s Internal Reference Input Input frequency 10MHz 50Hz Level range Compatible with TTL Level range 50Ω typical value DC coupling Trigger input Input level Compatible with TTL Slop Rising or falling optional Pulse width 100ns Input impedance 10kΩ DC coupling Response time 1μs typical value Trigger output Level Compatible with TTL Pulse widt...

Page 158: ... resolution Communication interface USB Host USB Device LAN Supply voltage 100 240 VAC 45 440Hz CAT II Power dissipation Less than 50W Fuse wire 2A T class 250V Environment Temperature range operating 10 40 Non operating 20 60 Cooling method Forced cooling by fan Humidity range 35 以下 90 relative humidity 35 40 60 relative humidity Altitude Operating below 2000 meter Non operating below 15000 meter...

Page 159: ...UTG9000T Series Function Abitrary Waveform Generator 159 Appendix C Accessories List Model UTG9000T Standard A power line up to local standard A USB data line Four BNC cables 1meter Optional ...

Page 160: ...Cut off the power and then clean with soft cloth that is wet but not dripping wipe floating dust off the exterior of instrument with mild detergent or clear water don t use chemical medicine or detergent containing benzene methylbenzene dimethylbenzene acetone and other potent substances Be careful scratch LCD protection screen when cleaning the instrument Protect the instrument away from any corr...

Page 161: ...UTG9000T Series Function Abitrary Waveform Generator 161 This user manual may be revised without prior notice ...

Page 162: ...UTG9000T Series Function Abitrary Waveform Generator 162 ...

Reviews: