19
10.5.2.
Removing the Internal Electrode
Handle the internal electrode with clean gloves. If the internal
electrode is stained with oil or contaminant, the pressure indication
may be affected by gas discharge.
Wear clean gloves.
10.5.2.1.
Removing the cap screw
Before the cap screw is removed
After the cap screw is removed
Turn to remove.
Fig. 10-4 Removing the cap screw
10.5.2.2.
Removing the anode
This can be a cause of leak. Do not bend the anode pin or give an
impact to the black glass portion.
Impact to anode pin and glass part
Push the metal portion on the outer periphery of the anode gradually to make the anode
aslant. Take this step from each of 4 directions.
Fig. 10-5 Removing the cap screw
Glass
①
②
④
③
Summary of Contents for G-Tran BMR2