background image

PAGE 27 OF 33 

1219 IH-8105

BITS 2, 1 AND 0

Weight Description

2

1

0

0

0

0

lb or kg, selected by Status Byte B, 

bit 4

0

0

1

grams (g)

0

1

0

metric tons (t)

0

1

1

ounces (oz)

1

0

0

not used

1

0

1

not used

1

1

1

tons (ton)

1

1

1

no units

BIT 3

Print Request = 1

BIT 4

Expand Data x 10 = 1, Normal = 0

BIT 5

Always = 1

BIT 6

Always = 0

COMMAND

FUNCTION

IP

Immediate Print of displayed weight (stable or unstable).

P

Print displayed weight (stable or unstable).

CP

Continuous Print.

SP

Print on Stability.

xS

0S: Turn off "Stable Only" menu item and allow unstable print. 1S: Turn on “Stable Only” menu item and only print stable print.    

xP

Interval Print x = Print Interval (1-50,000 sec), 0P turns auto print OFF.

Z

Same as pressing Zero Key.

T

Same as pressing Tare Key.

xT

Download Tare value in grams (positive values only).  

Sending 0T clears tare (if allowed).

PU

Print current unit: g, kg, lb, oz, lb:oz

xU

Set scale to unit x: 1=kg, 2=lb, 3=g, 4=oz, 5=lb:oz

xM

Set scale to mode X. 1=Weighing, 2=Counting, 3=Check, 4=Percent, 5=Dynamic. M will scroll to next enabled mode.

PSN

Print Serial Number.

CU xxx

Set Under Limit (only in Check mode) where ‘xxx’ is the value under current unit.

CO xxx

Set Over Limit (only in Check mode) where ‘xxx’ is the value under current unit.

x#

Set Counting APW (x) in grams. (only in Counting or Check Counting mode, must have APW stored).

P#

Print Counting or Check Counting mode APW.

x%

Set Percent mode reference weight (x) in grams (must have a weight stored).

P%

Print Percent mode reference weight.

PV

Version: print name, software revision and LFT ON (if LFT is set ON).

H x "text"

Enter String content, x = String number (1-10), "text" = string text up to 40 alphanumeric characters.

\EscR

Global reset to reset all menu settings to the original factory defaults.

STATUS BITS FUNCTION
Bit 0

Gross = 0, Net = 1

Bit 1

Sign, Positive = 0, Negative = 1

Bit 2

Out of Range = 1 (Over Capacity or Under Zero)

Bit 3

Motion = 1, Stable = 0

Bit 4

lb = 0, kg = 1, (see also Status Byte C, bits 0, 1, 2)

Bit 5

Always = 1

Bit 6

Zero Not Captured after power-up

Status Byte B Bit Definitions

OHAUS Commands

Status Byte C Bit Definitions

The Indicator supports both MT-SICS and OHAUS commands. Commands listed in the following tables will be 

acknowledged by the indicator. To use the MT-SICS commands, send the command PSI. To return to the OHAUS 

commands, send the command POH. SICS commands can also be active in the menu setup.

SERIAL COMMUNICATION AND PRINTER SETUP CONTINUED

Summary of Contents for OHAUS DEFENDER 5000

Page 1: ...its Print Short press Send the current display value to the RS 232 port or option when properly enabled Long Press Change the current weighing unit Press and hold the key to scroll through the list of...

Page 2: ...1 a Switch between three input modes numeric lowercase and uppercase input OVERVIEW OF CONTROLS SETUP 1 Remove shipping spacers and place the weighing pan securely over the rubber load pads on the top...

Page 3: ...ate Edit Template Edit String Reset WiFi Bluetooth WiFi MAC Address Network Port DHCP IP Address Gateway DNS Subnet Mask Alternate Command Reset Bluetooth Device Name Print Setup Assignment Print Opti...

Page 4: ...FT set to ON 4 Allow the scale to warm up for approximately five minutes after stabilizing to room temperature 5 To abort calibration press the softkey corresponding to the X icon anytime during the c...

Page 5: ...to enter the Calibration submenu Scroll to Linearity Calibration using the softkey corresponding to the icon See Figure 14 3 Press the softkey corresponding to the icon to initiate Linearity Calibrati...

Page 6: ...10 19 02 7 7 6 6 5 5 4 4 3 3 2 19 02 20 45 8 7 7 6 6 5 5 4 4 3 3 20 45 22 22 8 8 7 7 6 6 5 5 4 4 3 22 22 23 54 9 8 8 7 7 6 6 5 5 4 4 23 54 25 21 9 9 8 8 7 7 6 6 5 5 4 25 21 26 45 10 9 9 8 8 7 7 6 6 5...

Page 7: ...6 26 25 85 45 90 00 31 30 30 29 29 28 28 27 27 26 26 OPERATION CONTINUED TABLE OF GEO VALUES CONTINUED SETUP MENU When the Indicator connects to a scale base for the first time enter this menu to set...

Page 8: ...tion for the second weighing range of the scale If Single interval is enabled this parameter will not be shown NOTE Graduation settings are limited to values from Capacity divided by 600 to Capacity d...

Page 9: ...intain zero until a change of one division per second is exceeded 3d Display will maintain zero until a change of three divisions per second is exceeded AUTO DIM Set the display backlight functionalit...

Page 10: ...late Discrete Output 1 Off Overload Underload Zero Discrete Output 2 Off Overload Underload Zero Discrete Output 3 Off Overload Underload Zero Discrete Output 4 Off Overload Underload Zero Counting En...

Page 11: ...actor multiplier 3 divide the Factor by 1 000 1x10 3 2 divide the Factor by 100 1x10 2 1 divide the Factor by 10 1x10 1 0 multiply the Factor by 1 1x100 1 multiply the Factor by 10 1x101 2 multiply th...

Page 12: ...ss the button Mode to select an activated application The current application will be shown in the upper left corner of the home screen The Indicator incorporates the following applications WEIGHING U...

Page 13: ...esults 1 To view accumulation results press the softkey corresponding to the icon The Accumulate Result screen is displayed See Figure 31 2 To clear accumulation results press the Off On CLR button Se...

Page 14: ...umber of pieces SET THE AVERAGE PIECE WEIGHT APW NOTE It is recommended that the APW is larger than 1d If APW is between 0 05d and 1d a warning message will be displayed and the information line will...

Page 15: ...late Off Automatic Manual Off The icon does not display Manual The softkey displays User can press the key to accumulate the stable weight Automatic The icon displays The weight will be accumulated au...

Page 16: ...Figure 54 CHECK Check is used to compare the weight or pieces of a sample against target limits Press the Mode Library button until the icon corresponding to Check is displayed on the screen Two diffe...

Page 17: ...and Over Accept Under Over Off No beep Under and Over Accept Under Over Beep when reaching the selected check point CHECK COUNTING 1 Press the button and select Check Mode to Check Counting See Figure...

Page 18: ...e stable weight Automatic The icon displays The weight will be accumulated automatically NOTE The load to be accumulated has to be 10d The next accumulation can only start once the pan is cleared When...

Page 19: ...e Figure 65 2 Press the softkey corresponding to the icon for confirmation See Figure 66 Method 2 1 Press the softkey corresponding to the icon to enter the submenu for setting reference weight See Fi...

Page 20: ...list and highlight Application Mode See Figure 72 2 Enter this submenu by pressing the button corresponding to the icon See Figure 72 3 In the Application Mode menu enter the Percent submenu See Figu...

Page 21: ...e placed on the pan first After the terminal has done taring automatically place load on the pan to start dynamic weighing Start Weight 3 000 with current unit If the operation type is Semi Automatic...

Page 22: ...s setup will only work when the I O Option board has been installed The option I O board provides two isolated inputs and four dry contact normally open relay outputs which can be used for simple pro...

Page 23: ...table Only Off On LFT Force On Auto On Stable Mode Load Load and Zero Auto On Accept Interval Time 1 50 000 MT Continuous OH Continuous SICS Reference Balance Select Template Simple Custom 1 Custom 2...

Page 24: ...lected and weighing mode is Check values will be printed when weight is accepted ACCEPT Printing occurs each time the display is within the Checkweigh accept range and stability criteria are met INTER...

Page 25: ...minate a template an End of Template field must be included All fields after the End of Template field will be ignored Figure 87 Figure 88 ITEM LENGTH ITEM LENGTH 3 spaces 3 String 1 Not fixed up to 4...

Page 26: ...structure 3 Displayed weight Either gross or net weight Six digits no decimal point or sign Insignificant leading zeroes are replaced with spaces 4 Tare weight Six digits of tare weight data No decim...

Page 27: ...re xxx is the value under current unit CO xxx Set Over Limit only in Check mode where xxx is the value under current unit x Set Counting APW x in grams only in Counting or Check Counting mode must hav...

Page 28: ...nt weight value immediately SIRU Send weight value in the current unit immediately and repeat SIU Send weight value in the current unit immediately SNR Send stable weight value and repeat after every...

Page 29: ...ify communication by pressing the print button NOTE If HyperTerminal is set up properly the value on the display will be shown in the window CONNECTING TO A SERIAL PRINTER Connect the cable supplied w...

Page 30: ...used to separate the other fields The Weight1 field is four right justified characters If the value is negative the character is located at the immediate left of the most significant digit The Unit1...

Page 31: ...k on the terminal BATTERY OPERATION During operation battery icon indicates charge level remaining The battery icon is located above weighing unit on indicator display See Figure 94 NOTE When battery...

Page 32: ...verification procedure SEALING The local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings Refer to the illustrations below fo...

Page 33: ...ected menu to off in the Lock menu Lockout Switch on the circuit board may need to be set to the off position Error 8 1 Weight reading exceeds Power On Zero limit Remove load from scale Recalibrate sc...

Reviews: