PAGE 11 OF 33
1219 IH-8105
OPERATION CONTINUED
WEIGHING UNIT
Enter this menu to activate the desired units. Default
settings are
bold
.
NOTE: Due to national laws, the indicator may
not include some of the units of measure listed.
If the security switch is set to ON, the units are
locked at their current setting.
Choose desired unit, and set the status as listed below:
• GRAM (G)
• KILOGRAM (KG)
• POUND (LB)
• OUNCE (OZ)
• POUND: OUNCE (LB: OZ)
• TONNE (METRIC TONNE)
• TON (SHORT TON)
Status:
Off
= Disabled
On = Enabled
CUSTOM UNIT (C)
Use the custom unit to display weight in an alternative
unit of measure. The custom unit is defined using a
conversion factor, where the conversion factor is the
number of custom units per gram expressed in scientific
notation (Factor x 10^Exponent).
Factor
Set the conversion factor using the numeric keypad.
Settings of 0.1000000 to 1.9999999 are available.
The default setting is 1.0.
Exponent
Set the factor multiplier.
-3 = divide the Factor by 1,000 (1x10
-3
)
-2 = divide the Factor by 100 (1x10
-2
)
-1 = divide the Factor by 10 (1x10
-1
)
0 = multiply the Factor by 1 (1x10
0
)
1 = multiply the Factor by 10 (1x10
1
)
2 = multiply the Factor by 100 (1x10
2
)
Least Significant Digit
Set the graduation.
Settings of 0.5, 1, 2, 5, 10 and 100 are available.
The custom unit’s name can be customized up to three
characters.
NOTE: Custom Unit is locked at off position
when the Security Switch is set to the locked
position. Custom Unit is not available when
Range is set to Dual interval.
Set the status.
Off
= Disabled
On = Enabled
GLP/GMP MENU
Enter this menu to set the Good Laboratory Practice
(GLP) or Good Manufacturing Practice (GMP) data.
Date Format
Set the date format.
MM/DD/YYYY = Month/Day/Year
DD/MM/YYYY = Day/Month/Year
YYYY/MM/DD = Year/Month/Day
DATE
Set the date.
00 to 9999 = year position
01 to 12 = month position
01 to 31 = day position
Refer to Menu Navigation (See Page 4) to enter settings.
TIME FORMAT
Set the time format.
24 hr = 24-hour format.
12 hr = 12-hour format.
TIME
Set the time.
24-hour format
00 to 23 = hour position
00 to 59 = minute position