KANGAROO
e
Pump
™
Enteral Feeding Pump
figure . Running screen with the >
HRs of Pump set Usage indicator in upper
left of the screen.
leD array
The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially in darkened
rooms.
A green light status indicates normal pump operation.
A yellow light status indicates an informational or warning situation for a low battery condition, the pump in
HOLDING
mode, or one
of the following errors:
HolD eRRoR
,
feeD eRRoR
,
floW eRRoR
, and
flUsH eRRoR
.
A red light status indicates a critical pump error. The screen will show one of the following error messages:
sysTeM eRRoR
,
PUMP
seT DIsloDGeD
, or
RoToR eRRoR
.
section X — specifications and symbols
specifications
Ul listing
Medical electrical equipment
KANGAROO ePump™ Enteral Feeding Pump
(1) Classified with respect to electrical shock, fire, and mechanical hazards
in accordance with UL60601-1.
(2) Classified with respect to electrical shock, fire, mechanical and other
specified hazards in accordance with CAN/CSA C22.2 No. 601.1
Type Infusion Device
Volumetric
Pumping Mechanism
Rotary Peristaltic
Pump sets
KANGAROO ePump MISTIC Feed-Only Set or Feed & Flush Set
feeding formula Delivery Rate
1-400 mL/hr in 1 mL increments