Press
4
flUsH InTeRVal
in Figure 12A to define the time interval between starts of flushing. Press the
4
buttons in Figure 12C to
program the time interval from 1 to 24 hours in increments of 1 hour. Press
4
enTeR
(top button) to exit the menu.
The pump will automatically limit flushing settings which exceed pump capabilities to deliver.
CaUTIon:
The KANGAROO ePump enteral feeding pump flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use care when
programming the flush volume so that it matches the patient’s need.
noTe:
The pump will automatically flush an additional 25 mL of water after completing the programmed delivery of formula, as
long as at least one flush had occurred during formula delivery.
figure a.
ADJUST FLUSH
menu, continuous mode.
figure b. setting the flush
Volume.
figure C. setting the flush Time
Interval.
Intermittent Mode — adjust bolus
(Not Available in EZMODE)
Intermittent Mode feeding
For
InTeRMITTenT
feeding mode, the
aDJUsT bolUs
option is available in the opening menus (Figure 7C for a Feed-Only set or
Figure 7D for a Feed & Flush set). Select this option to program the feeding parameters from the
ADJUST BOLUS
screen, Figure 13A.
Press
4
bolUs RaTe
in Figure 13A to define the rate of delivery.
Press the
4
buttons in Figure 13B to program the delivery rate from 1 to 400 in increments of 1 mL/hr. Press
4
enTeR
(top button)
to exit the menu.
Press
4
nUMbeR of bolUses
in Figure 13A to define the total number of boluses to be delivered.
Press the
4
buttons in Figure 13C to program the number of boluses, up to a maximum of 99.
Press
4
enTeR
(top button) to exit the menu. The number of boluses can be set as zero, in which case the pump will run until the
supply is exhausted.
Press
4
VolUMe PeR bolUs
in Figure 13A to define the Volume to be delivered in each bolus.
Press the
4
buttons in Figure 13D to program the volume per bolus from 1 to 3000 mL in increments of 1 mL.
Press
4
enTeR
(top button) to exit the menu.
Press
4
bolUs InTeRVal
in Figure 13A to define the time interval between starts of bolus deliveries.
Press the
4
buttons in Figure 13E to program the time interval between starts of bolus deliveries from 1 to 24 hours in
increments of 1 hour. Press
4
enTeR
(top button) to exit the menu.
The pump will automatically limit bolus settings which exceed pump delivery capabilities. For example, a bolus delivery volume of
200 mL cannot be made once per hour if a rate of 150 mL/hr is already programmed. In this case, the maximum volume allowed
would be 150 mL.