![Tyco kENDALL KANGAROO ePump Operation And Service Manual Download Page 24](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-epump/kendall-kangaroo-epump_operation-and-service-manual_1186010024.webp)
KANGAROO
e
Pump
™
0
Enteral Feeding Pump
flush now
(Not Available in EZMODE)
During the
RUNNING
mode, the pump can be diverted to immediately perform a flush (flush on demand) by pressing
4
FLUSH
NOW
in the
RUNNING
menu, input a desired flush volume, then press
4
enTeR
. See Figure 15A. When
4
flUsH noW
is pressed,
the screen will prompt for the volume to be immediately flushed, and will show the previously programmed flush volume as default
but may be adjusted to the users desired volume. The screen will appear as the standard Flush input screen as shown in Figure 12B
(or 14B). Setting the
flUsH noW
flushing volume is temporary and will not change the main/periodic flushing volume previously
programmed. If
4
flUsH noW
is pressed by accident; the user may either press the cancel button or allow the screen to return to
the
RUNNING
screen after 10 seconds, if there is no input for the flush volume.
If stopping a
flUsH noW
before the set volume has been delivered is desired, press
4
flUsH noW
and enter a volume of zero and
press
4
enTeR
. The
flUsH noW
will stop immediately and the pump will return to the
RUNNING
mode.
More options
Press
4
MoRe
in one of the opening menus, Figures 5A-5B or Figures 7A-7D, to access the
MORE OPTIONS
screen, Figure 19.
figure .
MORE
OPTIONS
menu.
buzzer alarm Volume
Press
4
bUZZeR
in the
MORE OPTIONS
menu to access the
BUZZER
screen and increase or decrease the alarm buzzer volume. The
alarm will be audible as the volume setting is changed. This sets the volume level for any warning or error that may occur with the
pump. Critical errors of System Error and Pump Set Dislodged will always use the maximum volume setting.
History
(Not Available in EZMODE)
Press
4
HIsToRy
in the
MORE OPTIONS
menu to access the
HISTORY
screen. Press the
4
buttons in Figure 20 to set the time history
of interest. The totals of
feeD ml
and
flUsH ml
will be shown for the requested history time.
WaRnInG:
A maximum of 72 hours of past history can be retrieved, excluding data from the current hour. NOTE: The 72-hour
history includes time when the pump is powered OFF.