![Tyco kENDALL KANGAROO ePump Operation And Service Manual Download Page 31](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-epump/kendall-kangaroo-epump_operation-and-service-manual_1186010031.webp)
KANGAROO
e
Pump
™
Enteral Feeding Pump
system error
(Fixed Audible Volume)
The system error screen is the most general form of error, Figure 25. The only way to exit from a System Error is to power down. No
ConTInUe
option is allowed because of the severity of the error. An error number is also displayed on the screen, for reference
purposes. This number should be reported when calling the Customer Service Support Line.
error # Related sub-system
0
Internal timer stopped.
1
Flash Chip missing OR version incompatible.
2
Too many graphic screens in memory.
4
Never got a ADC ISR for motor current.
5
Never got a ADC ISR for battery level.
6
Never got a ADC ISR for battery temperature.
7
Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag-Empty checks.
8
Never got a ADC ISR for Ultrasonic during Downstream Occlusion.
9
RS232 Communications Transmission Error.
10
Flash Write Error.
11
Flash Read Error.
12
General Stopcock Time Out.
13
Time Out for Buzzer Test
16
Flash Screen Read Error.
19
Flash Languages Read Error.
20
Could not find Stop Cock Flush Position during prime routine.
21
Could not find Stop Cock Feed Position during prime routine.
22
Motor Error during Tubing AutoLoad.
23
Motor Failsafe Time Out Error. Possible damaged gearbox encoder or flex cable.
25
Display Lockup Time Out Error.
33
Motor Control Failure.
figure . General system
error screen. see Customer service