KANGAROO
e
Pump
™
Enteral Feeding Pump
figure . Installation of battery Pack.
WaRnInG:
USE OF A BATTERY PACK NOT SUPPLIED BY TYCO HEALTHCARE GROUP LP CAN BE DANGEROUS AND VOIDS ALL
WARRANTIES AND PERFORMANCE SPECIFICATIONS.
The battery will charge continuously whenever the pump is plugged into a wall outlet. Eight hours of charging is required to fully
recharge the Battery Pack.
section IX — error/Warning/Informational screens and Troubleshooting
For operational errors, the KANGAROO ePump enteral feeding pump display will show the error indicator icon, with specific
information regarding the error condition, and will sound the alarm.
The following conditions will trigger error conditions and will activate audible and visible alarms:
• Feed or Flush containers are empty
• Feed or Flush tube becomes occluded between bag and pump
• Tube becomes occluded between pump and patient
• Battery low (alarm beeps continuously, about 15 minutes before shut-off)
• Unit is left in HOLD mode longer than 10 minutes without input
• Pump Set tubing is improperly loaded around the rotor
• System errors, see below
• Feeding complete
• The MISTIC connector is removed during
AUTOPRIMING
or
RUNNING
The Error, Warning, and Information Screens are described below: