CaUTIon: avoid exposing power cord to excess moisture, as this can lead to an electrical shock or fire hazard.
Directions for Cleaning Rotor assembly
• Refer to General Cleaning Directions before starting.
• Open the blue door enclosing the Pump Set loading area.
• Loosen rotor set screw with 5/64” (2 mm) allen wrench and gently pull rotor off shaft. After removing rotor, avoid getting any
moisture in the rotor shaft opening.
• Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.
• Be sure all parts of rotor are completely dry before putting it back onto shaft.
• To replace rotor, align set screw on hub of rotor with the flattened section of the output shaft.
• Push rotor into place and tighten set screw (Do not over tighten).
Preventative Maintenance
This pump must be periodically re-certified to assure proper functioning and safety. The recommended service interval is at least
once per year. Re-certification may be done at the user’s Biomedical Engineering Department, an outside service, or by Tyco
Healthcare Group LP Factory Service. To arrange for Tyco Healthcare Group LP Factory Service in the US, call 1-800-448-0190. In
Canada, call 1-877-664-TYCO(8926). Outside of the US and Canada, please contact your local Customer Service for more information.
If a pump malfunctions, please contact your THGLP Representative or call Customer Service for instruction.
section VIII — battery Replacement
See Figure 24 and do the following to replace the Battery Pack:
• Power down the unit.
• Loosen the Battery Door screw, on back of pump.
• Open Battery Door and disconnect wire harness. A small straight-bladed screw driver may be needed to disconnect the connector
tab.
• Slide Battery Pack out of compartment.
• Slide a new Battery Pack into compartment with wire harness extending out.
• Align connector with red wire at upper right.
• Push connector in until properly latched.
• Tuck wires into slot between connector and battery to insure wires are not pinched. Re-tighten Battery Door screw.
note:
Replacement Battery Packs are available from Tyco Healthcare Group LP (See Section XIII - Service Part Numbers).