![Tyco kENDALL KANGAROO ePump Operation And Service Manual Download Page 16](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-epump/kendall-kangaroo-epump_operation-and-service-manual_1186010016.webp)
KANGAROO
e
Pump
™
Enteral Feeding Pump
Press
4
sToP
to cancel
aUTo PRIMe
.
When auto-priming is completed, the status line will show
AUTO PRIME COMPLETE
and will no longer show the
aUTo PRIMe
option. Be sure that the line(s) are fully primed. If not, use the hold-to-prime option(s) to complete the priming, as described below.
Hold-to-Prime
The hold-to-prime options allow for precise interactive control of the priming process.
feed-only Pump sets
For Feed-Only Pump Sets, the screen will appear as in Figure 8A. Press and hold
4
HolD To PRIMe feeD
until the feed line has
been primed down to the stepped end connector at the end of the Pump Set.
feed & flush Pump sets
For Feed & Flush Pump Sets, the display will appear as in Figure 8B. First press and hold
4
HolD To PRIMe flUsH
to prime the
flushing line, and then press and hold
4
HolD To PRIMe feeD
to prime the feed line down to the stepped end connector at the end
of the Pump Set.
If a Feed & Flush Pump Set is used, remember to first press
4
HolD To PRIMe flUsH
until the flushing fluid has reached the valve,
and then press
4
HolD To PRIMe feeD
until the feeding fluid has reached past the valve and down to the stepped end connector.
If the feeding line is primed first, subsequent priming of the flushing line will force the air between the flush solution bag and valve
into the main line, which already has been primed with enteral formula.
Re-priming after bag empty
A Pump Set bag that has been emptied will trigger the
feeD eRRoR
screen. In this condition the Pump Set bag can be refilled to
continue the feeding, but only after the Pump Set has been re-primed.
To re-prime the Pump Set, do the following:
• Disconnect the feeding line from the patient
• Refill the bag
• Press
4
ConTInUe
to begin the pump
RUnnInG
• Press
4
HolD
• Press
4
aDJUsT
seTTInGs
• Press
4
PRIMe PUMP
• Press
4
HolD To PRIMe
to prime the line(s)
Pump Sets should not be reused after 24 hours of initial usage.
selecting feed Mode
eZMoDe option
The
eZMoDe
option is a limited,
ConTInUoUs MoDe
only setting of ePump. It feeds non-stop (no VTBD option) at a rate