KANGAROO
e
Pump
™
Enteral Feeding Pump
Select
4
ConTInUoUs
/
InTeRMITTenT
under the
MORE OPTIONS
menu, Figure 10A. Then press
4
ConTInUoUs MoDe
or
4
InTeRMITTenT MoDe
from the
seleCT MoDe
menu, Figure 10B. Press
4
Done
to exit.
figure 0a.
MORE OPTIONS
menu.
figure 0b.
SELECT MODE
menu for
continuous or intermittent mode
selection.
Continuous Mode — adjust feed
Continuous Mode feeding
For
ConTInUoUs
feeding mode, the
aDJUsT feeD
option is available in the opening menus (Figure 7A for Feed-Only sets or Figure
7B for Feed & Flush sets). Select this option to program the feeding parameters from the
aDJUsT feeD
screen, Figure 11A.
Press
4
feeD RaTe
in Figure 11A to define the rate of delivery. Press the
4
buttons in Figure 11B to program the rate from 1 to 400
in increments of 1 mL/hr. Press
4
enTeR
(top button) to exit the menu.
Press
4
feeD VTbD
in Figure 11A to define the Volume To Be Delivered. Press the
4
buttons in Figure 11C to program the volume
from 1 to 3000 mL in increments of 1 mL. Press
4
enTeR
(top button) to exit the menu. If the
feeD VTbD
is set to zero, the pump
will run until the supply is exhausted.
noTe:
When a
feeD VTbD
is set, the remaining volume to be delivered (remaining VTBD) will be displayed next to the
W
icon on
the lower right-hand portion of the screen (Figure 11A). If the
feeD VTbD
is reprogrammed during a feeding, from
HOLDING
mode,
the pump will restart the feeding and deliver the entire volume of the new VTBD setting.
figure a.
ADJUST FEED
menu.
figure b. setting the feed Rate. figure C. setting the feed
Volume to be Delivered.
Continuous Mode flushing
(Not Available in EZMODE)
For
ConTInUoUs
feeding mode with a Feed & Flush Set loaded, the opening menu will appear as shown in Figure 7B.
Press
4
aDJUsT flUsH
in Figure 7B to program the flushing parameters from the
ADJUST FLUSH
screen, shown in Figure 12A.
Press
4
flUsH VolUMe
in Figure 12A to define the volume per flush occurrence. Press the
4
buttons in Figure 12B to program the
flush volume from 10 to 500 in increments of 1 mL. Press
4
enTeR
(top button) to exit the menu.