![Tyco kENDALL KANGAROO ePump Operation And Service Manual Download Page 17](http://html1.mh-extra.com/html/tyco/kendall-kangaroo-epump/kendall-kangaroo-epump_operation-and-service-manual_1186010017.webp)
KANGAROO
e
Pump
™
Enteral Feeding Pump
0-400mL/hr as defined by the user until manually stopped. It has no
bolUs
,
flUsHInG
,
RUn MoDe sCReen loCK
, or
aUToResUMe
capability and displays no
HIsToRy
information.
Turning on
eZMoDe
will automatically force the unit out of the
InTeRMITTenT (bolUs feeD)
mode if set. Although the
Continuous Mode feed VTBD will be set to 0 while
eZMoDe
is enabled, the previous value of the
feed VTbD
will be stored in flash
memory and recalled when “
eZMoDe
” is deselected. The user can change the Feed Rate by normal means. All other user settings,
including those not compatible with the
eZMoDe
(
bolUs VTbD
,
nUMbeR of bolUses
,
sUPeR bolUs
, etc), that were set
before the pump was put into the
eZMoDe
will be ignored and invisible while in
eZMoDe
but remain stored in the flash memory
for when
eZMoDe
is deselected.
eZMode
essentially has no effect on
CleaR seTTInGs
; pressing
CleaR seTTInGs
while
eZMoDe
is enabled will clear all Feed and Flush settings and totalizers as normal, not just those that apply to
eZMoDe
. Selecting
the
CleaR seTTInGs
option on power up will not clear
biotech options
such as
eZMoDe
or
loCK seTTInGs
.
If the user selects the
aDJUsT feeD
option from either the
loaD a seT
or
seT loaDeD
screens, control will pass directly to and
from the Enter Feed Rate screen and bypass the
aDJUsT feeD
screen since without the ability to adjust VTBD in
eZMode
, the Adjust
Feed screen is redundant.
selecting eZMoDe
eZMoDe
may be selected through the
biotech options
button which may be accessed on the Biotech screen (see page 21). The
eZMoDe
setting will be stored in the flash memory and saved when the unit is shut off.
noTe:
If the user inserts a Feed/Flush set while the
eZMoDe
is enabled,
eZMoDe
will be cancelled and the unit will immediately
revert back to the normal mode. If
eZMoDe
is desired, the user must then re-enable
eZMoDe
through the
biotech options
screen.
When enabled and not in
CeRTIfICaTIon MoDe
, a small
eZ
icon will be visible on the bottom of the screen near the battery icon.
However, it will not appear in the
PoWeR DoWn
,
sysTeM eRRoR
, or
loW baTTeRy sCReens
.
figure .
selecting Continuous or Intermittent feed Mode
The KANGAROO ePump enteral feeding pump can be programmed to feed continuously or intermittently.
The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has
been delivered or until the supply has been exhausted. The Intermittent feeding mode delivers boluses of enteral nutrition at
programmed time intervals. The bolus volume and feed rate are also programmed.
To select continuous feed mode or intermittent feed mode, press
4
MoRe
on the opening menu, shown in Figures 5A-5B or Figures
7A-7D, to access the
MORE OPTIONS
menu, Figure 10A.
The
MoRe
option is available either before a Pump Set has been loaded (Figures 5A-5B) or after the Pump Set is loaded (Figures 7A-
7D). See subsection “More Options” for information on the other options in the
MORE OPTIONS
menu.