Tyco kENDALL KANGAROO ePump Operation And Service Manual Download Page 17

KANGAROO 

e

Pump

Enteral Feeding Pump

0-400mL/hr as defined by the user until manually stopped.  It has no 

bolUs

flUsHInG

RUn MoDe sCReen loCK

, or 

aUToResUMe

 capability and displays no 

HIsToRy

 information.  

Turning on 

eZMoDe

 will automatically force the unit out of the 

InTeRMITTenT (bolUs feeD)

 mode if set.  Although the 

Continuous Mode feed VTBD will be set to 0 while 

eZMoDe

 is enabled, the previous value of the 

feed VTbD

 will be stored in flash 

memory and recalled when “

eZMoDe

” is deselected.  The user can change the Feed Rate by normal means.  All other user settings, 

including those not compatible with the 

eZMoDe

 (

bolUs VTbD

,

 nUMbeR of bolUses

sUPeR bolUs

, etc), that were set 

before the pump was put into the 

eZMoDe

 will be ignored and invisible while in 

eZMoDe

 but remain stored in the flash memory 

for when 

eZMoDe

 is deselected.  

eZMode

 essentially has no effect on

 CleaR seTTInGs

; pressing 

CleaR seTTInGs

 while

 

eZMoDe

 is enabled will clear all Feed and Flush settings and totalizers as normal, not just those that apply to

 eZMoDe

.  Selecting 

the 

CleaR seTTInGs

 option on power up will not clear 

biotech options

 such as 

eZMoDe

 or 

loCK seTTInGs

.

If the user selects the 

aDJUsT feeD

 option from either the 

loaD a seT

 or 

seT loaDeD

 screens, control will pass directly to and 

from the Enter Feed Rate screen and bypass the 

aDJUsT feeD

 screen since without the ability to adjust VTBD in 

eZMode

, the Adjust 

Feed screen is redundant.

selecting eZMoDe

eZMoDe

 may be selected through the 

biotech options

 button which may be accessed on the Biotech screen (see page 21).  The

 

eZMoDe 

setting will be stored in the flash memory and saved when the unit is shut off.  

noTe: 

If the user inserts a Feed/Flush set while the 

eZMoDe

 is enabled, 

eZMoDe

 will be cancelled and the unit will immediately 

revert back to the normal mode.  If 

eZMoDe

 is desired, the user must then re-enable 

eZMoDe 

through the 

biotech options

 

screen. 
When enabled and not in 

CeRTIfICaTIon MoDe

, a small 

eZ

 icon will be visible on the bottom of the screen near the battery icon.  

However, it will not appear in the 

PoWeR DoWn

sysTeM eRRoR

, or 

loW baTTeRy sCReens

.

figure .

selecting Continuous or Intermittent feed Mode

The KANGAROO ePump enteral feeding pump can be programmed to feed continuously or intermittently.  
The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has 

been delivered or until the supply has been exhausted.  The Intermittent feeding mode delivers boluses of enteral nutrition at 

programmed time intervals.  The bolus volume and feed rate are also programmed.
To select continuous feed mode or intermittent feed mode, press

4

MoRe

 on the opening menu, shown in Figures 5A-5B or Figures 

7A-7D, to access the 

MORE OPTIONS 

menu, Figure 10A.  

The 

MoRe

 option is available either before a Pump Set has been loaded (Figures 5A-5B) or after the Pump Set is loaded (Figures 7A-

7D).  See subsection “More Options” for information on the other options in the 

MORE OPTIONS

 menu.

Summary of Contents for kENDALL KANGAROO ePump

Page 1: ...gs og serviceh ndbog Pumpa pro enter ln v ivu Provozn a servisn p ru ka Enter lis t pl l pumpa Kezel si s karbantart si k zik nyv Pompa ywienia dojelitowego Instrukcja obs ugi i serwisowania Enteral B...

Page 2: ...entification 4 Section IV Initial Setup Attaching the A C Adapter Power Cord 5 Battery Setup 6 Attaching Pole Clamp 6 SectionV Instructions for Use Quick Start 7 General Startup 8 Placement Mounting 8...

Page 3: ...hing 16 Run 17 KTO 18 Lock Screen Feature 18 Hold 19 Flush Now 20 More Options 20 Buzzer AlarmVolume 20 History 20 Language Selection 21 Continuous Intermittent Mode 21 BioTech Menu 21 Accessing the B...

Page 4: ...s 31 Symbols on Pump and Accessories 34 Section XI Customer Service 35 Section XII Maintenance Side Door Enclosing Pump Set Loading Area 35 Battery Pack 36 Battery Door 36 Power Cord 36 Pole Clamp 36...

Page 5: ...otect Pump Set Integrated handle for easy transport Features MISTIC Magnetic Intelligent SetType Identification Connector Feed Only vs Feed Flush Set identification system to ensure match between the...

Page 6: ...views Menu Selections Menu Selection Buttons Graphic Display Feeding Set Cap Storage Area Front Door Back Battery Compartment Serial I O Port A C Adapter Port Pole Clamp Attachment Point Integrated Ha...

Page 7: ...ts 6 For re certification see SectionVI Re certification of Performance For other integrity checks consult with a qualified Biomedical technician or contact the manufacturer Section XI Customer Servic...

Page 8: ...fied as complying with the safety requirements for accessories used with the pump whereas the safety certification of the accessory has been performed in accordance to the appropriate IEC 60601 1 and...

Page 9: ...n of the pins and tab on the plug to properly orient the plug Figure 2A Alignment of A C Adapter plug with pump s adapter port EBQUFS 1PSU EBQUFS 1MVH 6TF DBSF BMJHOJOH BEBQUFS XJUI QVNQ DPOOFDUPS The...

Page 10: ...approximately 15 hours of service before needing to be recharged Attaching Pole Clamp The KANGAROO ePump enteral feeding pump can be attached to a vertical pole using the Pole Clamp included with the...

Page 11: ...straint clip is provided with the KANGAROO ePump enteral feeding pump Using the pole clamp mounting bolt follow the pole clamp assembly instruction to attach the cord clip Refer to Fig 3 Section V Ins...

Page 12: ...ng fluid has reached past the valve and down to the distal end connector 6 Use buttons4on left side of LCD to set the feeding parameters Set flushing parameters if a Feed Flush Set is loaded The conti...

Page 13: ...power up operations Should the language need to be changed after the initial power up it can be changed through the MORE OPTIONS menu see Figure 10A Keep or Clear Prior Pump Settings Immediately afte...

Page 14: ...elow 1 Open the blue transparent door that encloses the rotor and Pump Set loading area 2 Load the Pump Set by holding the finger tab on the valve and inserting the valve down into the slot on the lef...

Page 15: ...more interactive method using the hold to prime feature After a Pump Set has been loaded and the status line shows SETLOADED press4PRIME PUMP in the opening menu Figures 7A 7D to get one of the PRIME...

Page 16: ...a Feed Flush Pump Set is used remember to first press4HOLDTO PRIME FLUSH until the flushing fluid has reached the valve and then press4HOLDTO PRIME FEED until the feeding fluid has reached past the v...

Page 17: ...the Adjust Feed screen is redundant Selecting EZMODE EZMode may be selected through the Biotech Options button which may be accessed on the Biotech screen see page 21 The EZMode setting will be stored...

Page 18: ...rogram the volume from 1 to 3000 mL in increments of 1 mL Press4ENTER top button to exit the menu If the FEEDVTBD is set to zero the pump will run until the supply is exhausted NOTE When a FEEDVTBD is...

Page 19: ...from the ADJUSTBOLUS screen Figure 13A Press4BOLUS RATE in Figure 13A to define the rate of delivery Press the4buttons in Figure 13B to program the delivery rate from 1 to 400 in increments of 1 mL h...

Page 20: ...ing the time interval between starts of bolus Intermittent Mode Feeding Bolus Max Not Available in EZMODE The BOLUS MAX option can be used to deliver a bolus at a high rate similar to gravity feeding...

Page 21: ...settings which exceed pump capabilities to deliver CAUTION The KANGAROO ePump enteral feeding pump flushes at a rate of 1960 mL hr 32 7 mL minute Use care when programming the flush volume so that it...

Page 22: ...on will increment the KTO time by 5 minute increments up to 30 minutes see Figure 15B After 30 is reached each additional button press will increment in larger time steps up to a maximum of 240 minute...

Page 23: ...ST SETTINGS to return to one of the opening menus shown in Figures 7A 7D thus allowing any setting to be modified If the Pump Set has been removed after going into HOLD the opening menu will show LOAD...

Page 24: ...volume has been delivered is desired press4FLUSH NOW and enter a volume of zero and press4ENTER The FLUSH NOW will stop immediately and the pump will return to the RUNNING mode More Options Press4MOR...

Page 25: ...re information Press4DONE in the MORE OPTIONS menu to exit the menu BioTech Menu The BioTech screen will show technical information about the pump and has a lock feature to lock the Pump Settings agai...

Page 26: ...tton to indicate that the feature is active Pressing the4EZMODE button again will remove the check mark indicating that the feature is off If the EZMODE feature is turned on the EZMODE icon will appea...

Page 27: ...tifying the Accuracy of Pump Flow Rate Use the following procedure to check the flow rate accuracy Perform test at room temperature 22 C 2 C 72 F 3 F with a new Pump Set Fill a new Feed Only Pump set...

Page 28: ...follows CAUTION Disconnect pump from AC power source before cleaning After cleaning do not connect to AC power source until pump and power cord are thoroughly dry A mild detergent should be used for g...

Page 29: ...may be done at the user s Biomedical Engineering Department an outside service or byTyco Healthcare Group LP Factory Service To arrange forTyco Healthcare Group LP Factory Service in the US call 1 80...

Page 30: ...lay will show the error indicator icon with specific information regarding the error condition and will sound the alarm The following conditions will trigger error conditions and will activate audible...

Page 31: ...5 Never got a ADC ISR for battery level 6 Never got a ADC ISR for battery temperature 7 Never got a ADC ISR for Ultrasonic Upstream Occlusion or Bag Empty checks 8 Never got a ADC ISR for Ultrasonic d...

Page 32: ...olume The ROTOR ERROR screen appears during RUNNING or PRIMING when the pump has detected an unusual operating condition for the rotor See Figure 27 The ROTOR ERROR generally results from a problem wi...

Page 33: ...nd the patient Check the line to find the occlusion causing the blockage If the error cannot be resolved check the valve pocket in the Pump Set loading area for dampness or dirt which may be a cause f...

Page 34: ...hat the battery must be recharged immediately Feeding Complete Adjustable Audible Volume The FEEDING COMPLETE information screen appears after completion of the programmed feeding Figure 33 Press4Powe...

Page 35: ...d FLUSH ERROR A red light status indicates a critical pump error The screen will show one of the following error messages SYSTEM ERROR PUMP SET DISLODGED or ROTOR ERROR Section X Specifications and Sy...

Page 36: ...6 6 Width 16 3 cm 6 4 Depth 11 7 cm 4 6 Weight 1 1 kg 2 4 lbs 1 2 kg 2 7 lbs with pole clamp Material Housing ABS Polycarbonate blend Pump Set Door Polyester Polycarbonate blend Power Use AC adapter f...

Page 37: ...re 10 40 C 50 104 F 75 R H non condensing StorageTemperature 0 50 C 32 122 F 95 R H non condensing Type of Protection Against Electrical Shock Class II Internally powered Equipment Degree of Protectio...

Page 38: ...ree of protection against electrical shock Double Insulated Direct Current DC Drip Proof Degree of protection against ingress of fluids CE Mark European Community Certification Power Button Buttons fo...

Page 39: ...hcare Canada 98 6 Faichney Drive 7300Trans Canadian Highway Watertown NY 13601 Pointe Claire QC H9R 1C7 1 800 448 0190 or 315 788 5246 1 877 664 8926 or 514 695 1220 U K Service Center Outside of U S...

Page 40: ...that the battery be disconnected during this time The battery may need to be charged and discharged several times to restore optimal capacity Battery Door To replace Battery Door see Figure 36 and do...

Page 41: ...2 Figure 37 User Serviceable Parts 4 5 1 KANGAROO ePump US 382400 International 482400 2 Main Door Blue transparent door F31877WT 3 Battery Pack F010484WT 4 Battery Door with Screw F31878WT Power Cor...

Page 42: ...the defective part or product with a comparable part or product repair the defective part or product or if neither replacement nor repair is reasonably available refund to Customer the purchase price...

Page 43: ...4 3 2002 Complies Conducted Disturbance Immunity EN60601 1 2 IEC 61000 4 6 2001 Complies Power Frequency Magnetic Field Immunity EN60601 1 2 IEC 61000 4 8 2001 Complies Voltage dips and sags Immunity...

Page 44: ...uring process The KANGAROO ePump AccuSystem feeding set manufacturing process parameters are tightly controlled atTyco Healthcare Group LP to minimize variation and produce consistent and reliable pro...

Page 45: ...3 45 3 0 45 3 3 45 3 0 45 3 FFQ XBZ SPN 4VOMJHIU EEP RY 4JOHMF 6TF 05 6TF Z 3 5 4 4 61 45 3 0 P OPU VTF JG VOJU QBDLBHF JT PQFOFE PS EBNBHFE FFQ XBZ SPN FBU 4JOHMF 6TF 4FF DDPNQBOZJOH PDVNFOUT 5SBEFNB...

Page 46: ...spumpe Bruker og vedlikeholdsh ndbok Operation and Service Manual See Accompanying Documents 0123 Rx ONLY TM KANGAROO ePump Trademark of Tyco Healthcare Group LP or its affiliate SPRAY NINE is a trade...

Reviews: