Page 44 of 57
10.
OPERATING INSTRUCTIONS
It is recommended to perform a B&D test cycle at the beginning of each
working day.
10.1.
Filling the Mineral-Free Water Reservoir.
Attention: Only use water having the characteristics outlined in
section 4.10 Water Quality. Using Tap water will clog the
system and invalidate the manufacturer’s guarantee
.
1.
Open the door (3).
2.
Pour water, gently, into the front funnel (7) until it reaches the
required level
(4) on the sight gauge.
Please note that the sight gauge is divided into three sections. The bottom
red section (6) indicates that the water level in the reservoir is too low.
The middle blue section (4) shows that there is sufficient amount of
water. The upper red section (2) indicates that the water level in the
reservoir is too high. To assist the operator to detect the water level, the
sight gauge is equipped with a yellow float (5).
Caution!
Under no circumstance should water be filled above the top red
section on the sight gauge.
3.
If the reservoir is empty, it is recommended to add water directly into
the reservoir, as follows:
Remove the water reservoir cover to reveal the water reservoir (1).
Pour
water, having the characteristics outlined in section 4.10
Water Quality, into the
reservoir
through
the
opening
on
top
of the
autoclave
until
it
reaches
the blue level on the sight gauge (4).
Caution: Under no circumstance should water be filled above the
level of 1" below the safety valve holder (8).
In case more water is added accidentally above the top
red section
on the
sight gauge, decrease the water level by draining the reservoir before
starting a cycle (see para. 11.2).
7
5
1
2
6
3
4
8