Page 34 of 57
7.2.
Parameters
This directory enables the operator to see and change the cycle’s
parameters. Therefore it is necessary to choose the required cycle
before entering the "
MENU
" directory (see 10.6 Selecting a
program").
For seeing or changing the parameters proceed as follows:
The
CYCLE PARAMETERS
screen is displayed when entering the
"
PARAMETERS
" directory:
The following parameters of the current cycle are displayed:
1.
Ster Time: sterilization exposure time in minutes.
2.
Dry Time: drying time in minutes.
To select the desired parameter, use the
UP
and
DOWN
keys. When
the desired parameter is blinking press
START/STOP
.
The default SterTime value is the minimum allowable value. It is
possible to increase the time value up to a maximum of 99 minutes.
However, cycles of this length have not been tested and there is no
guaranty that they will be successful.
1.
To move the curser from one digit to another press the
START/STOP
key.
2.
Use the
UP
and
DOWN
keys to increase or decrease the current
digit.
3.
After changing the parameter to the required value, confirm the
new value. Do this by pressing
START/STOP
a few times until
SET
is blinking and then press
UP
or
DOWN
.
4.
If, before confirmation you decided not to change the parameter,
press
START/STOP
a few times until
EXIT
is blinking. Press
UP
or
DOWN
. The display will return to the previous menu and
the parameter will not be changed.
5.
If, by mistake, the curser skipped the
SET
and reached
EXIT
,
continue to press
START/STOP
until the curser returns to
SET.