5.2.
Displayed Error and Operational Messages
Error
Number
Error
Name
Message Description
Corrective Action
Err:001
Pt1 out
This error # and symbol are
displayed when the Chamber
Temperature sensor Pt1 is
disconnected or out of range.
Call the technician.
Err:002 Low
Temperature
This error # is displayed when
the temperature drops for more
than 1 second below the
sterilization temperature.
Check that the
autoclave is not
overloaded or check
for leakage from the
door.
If the problem
persists, call the
technician.
Err:003 High
Temperature
This error # is displayed in one
of the following cases:
During Pre-vacuum, Heating
or Sterilization Stages:
If Chamber Temperature is
greater than Limit Temperature
for more than 10 seconds.
Sterilization Stage:
If temperature rises 3°C (6°F)
above sterilization temperature
during the sterilization stage
for 10 seconds.
Check that the
autoclave is not
overloaded or
perform a new
cycle.
If the problem
persists, call the
technician.
Err:005 High Gen
Press
This error # is displayed when
the Generator pressure is
more than 40kPa (6psi) above
the required pressure.
Perform a new
cycle.
If the problem
persists, call the
technician.
Err:007 High
Pressure
This error # is displayed if
Chamber Pressure raises 20kPa
(3psi) above sterilization
pressure for 5 seconds during
the sterilization stage.
Perform a new
cycle.
If the problem
persists, call the
technician.
Err:008 Low
Pressure
This error # is displayed if
Chamber Pressure drops
bellow the sterilization
pressure for 5 seconds during
the sterilization stage.
Check for leakage
from the door.
Perform a new
cycle.
If the problem
persists, call the
technician.
Err:009 Time Error This error # is displayed if the
real time clock is faulty.
Call the technician.
Err:011
Door is
Open
This error # is displayed when
the door is open during the
cycle.
Open and close the
door, then run a
new cycle.
If the problem
persists, Call the
technician.