Page 42 of 57
11.
At least once a week use a biological spore test (Bacillus
Stearothermophilus) in any load to insure proper sterilization. (
Be aware
,
testing standards may vary). Always follow the spore test manufacturer’s
instructions.
12.
Make sure that all instruments remain apart during the sterilization cycle.
Surfaces that are hidden because items are covering other items will not
be exposed to the steam and will not be sterilized.
13.
Verify that packaging methods are in accordance with the good practice
approach and the packaging materials used are in agreement with
applicable standards.
14.
Empty canisters should be placed upsidedown in order to prevent the
accumulation of water.
15.
Do not overload the Sterilizer trays. Overloading will cause inadequate
sterilization & drying.
16.
Allow a distance of approximately 1" between trays to permit steam
circulation.
17.
Wrapped instruments should be placed in material which will allow
steam penetration and promote drying, such as autoclave bag, autoclave
paper, or muslin towels.
18.
Do not stack pouches. It is recommended that a pouch rack such as the
Tuttnauer
POUCH RACK
be used to insure proper steam penetration
and adequate drying. Surfaces that are hidden because the items are being
stacked will not be exposed to the steam and will not be sterilized.
19.
Tubing should be rinsed after cleaning. When placed in the tray, make
sure that both ends of the tubing are open and there are no sharp bends or
twists.
20.
Packs should be placed upright on the tray. They should not be touching
each other or the Chamber walls. There should be about 1” between
packs for proper steam circulation.