PRINTER OUTPUT
DESCRIPTION
Operator:
To be filled in manually by operator
Time:
15:01:48
Time sterilization cycle ended
Date:
04/29/2010
Date sterilization cycle ended.
CYCLE ENDED!
D 00:57:07
191.7
002.8
The time, temperature and pressure during drying.
*
*
Prints drying data every 3 minutes.
*
D 00:36:49
215.3
000.6
The time, temperature and pressure during drying.
-----------------------
E 00:36:32
220.1
001.6
The time, temperature and pressure during exhaust.
E 00:35:32
252.2
016.8
The time, temperature and pressure during exhaust.
-----------------------
S 00:35:32
252.2
016.8
The time, temperature and pressure during sterilization.
*
*
Prints sterilization data every 1 minute.
*
S 00:15:32
250.7
016.9
The time, temperature and pressure during sterilization.
----------------------
H 00:15:28
250.2
016.5
The time, temperature and pressure during heating.
H 00:12:28
158.4
014.7
The time, temperature and pressure during heating.
----------------------
V 00:12:28
156.6
014.7
The time, temperature and pressure during air removal.
*
Prints sterilization data every 3 minute or on upper and
lower peak (the shorter time)
*
*
V 00:00:00
108.0
009.9
The time, temperature and pressure during air removal.
Time
°F
PSI
Time, temperature and pressure (psig) columns
Dry Time
:020.0min
Drying time for selected program.
Ster Time
:020.0min
Sterilization time for selected program.
Ster Temp
: 250°F
Sterilization temperature in chamber for selected program
Cycle: WDry 121
Cycle name (with dry).
Load number: 0033
Cycle counter
Param. Ver: 0032
Version of parameters setting
Ver: ELARA 1.21
Software version.
Ser.Num: 02903190
Autoclave's serial number.
Time: 14:04:41
Time sterilization cycle started.
Date: 04/29/2010
Date sterilization cycle started.
POWER ON
The device is turned on
--------------------
Legend
V Vacuum
stage
H Heating
stage
S Sterilization
stage
E Exhaust
stage
D
Drying stage
T Test