Emergency opening of the door
If there is no power to the autoclave and the door is closed the lock will
be engaged. If you need to open the door, follow this procedure:
1.
Turn the power switch off.
2.
Wait at least one hour until the autoclave cools down.
3.
Verify that there is no pressure in the autoclave chamber by pulling
the ring of the chamber safety valve (see sec 11.5).
4.
Insert a pin (1) into the hole (2) in the handle, pressing down the
locking pin.
5.
Open the door.