![TriMedika TRI TEMP Instructions For Use Manual Download Page 8](http://html1.mh-extra.com/html/trimedika/tri-temp/tri-temp_instructions-for-use-manual_1162943008.webp)
TRITEMP™
Non-contact thermometer
EN
8
9
III. Safety Precautions
This thermometer is designed for use
by trained healthcare personnel only
and must be used in accordance with
the instructions in this manual. All
users must familiarise themselves with
the safety instructions and guidance
detailed in this manual before
operating the device.
Warnings and cautions can appear on
the thermometer, the packaging, the
shipping carton, or in the user manual.
Failure to understand and observe
warnings in this manual could lead to
patient injury and/or damage to the
thermometer which could affect the
accuracy of the device.
•
TRITEMP™
is a precision device
and must be handled carefully and
not used if dropped.
• The thermometer is not water proof
and must be stored in a clean dry
area and never submerged in water
or any other liquid.
• Do not expose the thermometer to
direct sunlight.
• Do not autoclave the thermometer.
• Do not point the thermometer at a
heat source before use as it could
raise the sensor temperature and
produce erroneous results.
• The thermometer should only be
used for the purposes described
in this manual.
• The thermometer should not be
exposed to electric shocks.
• Body temperature measurements
are determined for forehead
readings only and accuracy is not
guaranteed for other locations on
the body.
• The device must be used in ambient
temperatures ranging from 10°C
(50°F) to 40°C (104°F). Do not
expose to extremes of temperature
>55°C (131°F) or <-20°C (-4°F).
• Do not use at a relative humidity
higher than 85%.
• Do not touch the sensor with your
fingers and only clean sensor in
accordance with the instructions
in this manual.
• Always ensure the sensor is clean
and free from all dust, debris and
moisture before use to ensure
accurate readings.
• Ensure the thermometer is between
3-5cm (2 fingers width) from the
forehead before taking a reading.
Accuracy cannot be guaranteed
outside this recommended range.