![TriMedika TRI TEMP Instructions For Use Manual Download Page 19](http://html1.mh-extra.com/html/trimedika/tri-temp/tri-temp_instructions-for-use-manual_1162943019.webp)
TRITEMP™
Non-contact thermometer
EN
18
19
XIII. Change the
batteries
XIV. Troubleshooting and
Error messages
• Please refer to this guide to help
resolve any problems using the
device. If the problem persists
please contact your supplier or
TriMedika customer service.
• The device has Error messages that
will be displayed on the Display as
follows:
Lo
appears on the screen when the
BODY temperature measured is under
the measuring range or <32°C (89.6°F)
HI
appears on the screen when the
BODY temperature measured is above
the measuring range or >42.9°C
(109.2°F)
Reasons for Lo message
Action
Hair or perspiration on forehead
Remove hair or any other obstruction
prior to taking a temperature
Air draft or rapid change in
ambient temperature
Ensure there is no direct air flow in the
area of use.
Readings are too close together
Ensure there is 3-5 seconds between
temperature readings.
Thermometer is held too far away
from the forehead
Temperature must be taken 3-5cm
(2 finger widths) from forehead.
Measurement taken after cold
compress, taking medicine or bathing.
Ensure environment temperature is
between 10-40°C
Reasons for HI message
Action
When body temperature is above
42.9°C (109.2°F)
Make sure there is no source of hot air
in area of use. Seek immediate medical
attention.
Ambient temperature is above
40°C (104°F)
Ensure the operating temperature
is between 10°C to 40°C (50°F to104°F).
When surface temperature is
above 60°C (140°F)
Temperature reading exceeds
measurement range of 0°C to 60°C
(32°F to 140°F).