IX. How to take a 

measurement of patient

•  Allow the device to equilibrate to

  room temperature for 15-20 minutes

  before use. This is necessary to ensure  

  the sensor is at room temperature.

• 

Press the POWER button

 

(See Fig 1.

  Button Functions)

 

and release

 immediately

 to turn on the device.

  The screen will light Green and

  current settings will be displayed.

•  Check the screen settings are the  

  required parameters i.e. Body

  Temperature, °C and the speaker

  symbol is visible (for audio). If the

  settings are not those required 

(See  

  Section XI. Adjustment of settings) 

for  

  adjustment of the settings.

•  Ensure the site for taking

  temperature is exposed and is 

  free from hair, sweat, etc. 

•  Remove any covering (e.g. hat, dressing,  

  etc.) that would insulate the forehead  

  and prevent release of infra red.

•  Point the 

TRITEMP™

 straight

  onto the forehead (not at an angle)

  approximately 3-5cm (2 finger

  widths) from the skin and press the

  READ button. The device will ‘bleep’

  to indicate a temperature has been

  taken and the reading is displayed  

  on screen in 1 second. 
•  Screen colour indicator appears

  o GREEN

  

≤37.3°C (99.8°F)

    Normal range

  o ORANGE

  

37.4–37.9°C (100 -100.2°F) 

    Slight fever

  o RED

  

38°C (100.4°F)

    Febrile

•  The device has a HIGH temperature

  alarm which will sound if the Body

  temperature is 38°C (100.4°F) or

 above.
•  The device will automatically switch off  

  in 30 seconds to preserve the battery.

•   If you press the POWER button and

  hold until there is an audible bleep,

  °C will start flashing on the green

  screen. You are now in the MODE

  settings Menu. 

•   If you inadvertently enter the MODE

  settings menu, continue to press and

  release the POWER button a further

  4 times to scroll through the options

  and turn off the device. 

•   To restart the device, press and

  release the Power button to turn the

  Thermometer ON. 

Please refer to 

Section XI.

 Adjustment 

of 

MODE

 setting 

‘SETTING THE 

READ MODE’

X. How to take the 

measurement of an object

TriTemp™

Non-contact thermometer

8

IV. Symbols

ENG

WARNING

Meets essential requirements of European
Medical Directive 93/42/EEC

Consult operating instructions

WEEE Directive (2002/96/EC)

IEC 60529 Ingress Protection

Fragile

Keep Dry

Serial Number

IEC 60417-5034 Direct Current

IEC 60417-5333, Type BF applied part

Final TriTemp IFU Manual V3 - 0817 ENG-D.qxp_Layout 1  12/09/2017  12:24  Page 10

XIV Dépannage et messages d’erreur 

•  En cas de problèmes lors de l’utilisation
 

de cet appareil, veuillez consulter ce

 

guide.  Si le problème persiste, veuillez

 

contacter votre fournisseur ou votre

 

service client TriMedika.

•  L’appareil peut à l’occasion afficher les
 

messages d’erreur suivants :

Lo

 s’affiche à l’écran lorsque la température  

corporelle mesurée est en deça de la plage 
de mesure ou <32 °C (89,6 °F). 

HI

 s’affiche à l’écran lorsque la température  

corporelle mesurée est au-delà de la plage 
de mesure ou >42,9 °C (109,2 °F). 

65

TriTemp

TM

 

Non-contact thermometer     

 

F

Raisons pour lesquelles « Lo » apparaît 

Que faire

Cheveux ou transpiration sur le front 

Retirer les cheveux ou autre obstacle

 

avant de prendre la température

Courant d’air ou changements rapides de 

Veiller à ce qu’il n’y ait pas de souffle d’air

la température ambiante 

direct sur la zone d’utilisation.

Les prises de mesure sont trop 

Patienter 3 à 5 secondes entre les relevés

rapprochées 

de température.

Le thermomètre est trop loin du front 

La température doit être prise à une

 

distance de 3 à 5 cm (2 largeurs de doigt)

 

du front.

Mesure prise après application d’une 

La température ambiante doit être de 10 à

compresse froide, la prise de 

40 °C

médicaments ou un bain.

Raisons pour lesquelles « HI » apparaît  

Que faire

Lorsque la température du corps est 

Veiller à ce qu’il n’y ait pas de source d’air

supérieure à 42,9 °C (109,2 °F) 

chaud en utilisation.  Consulter

 

immédiatement un médecin.

La température ambiante est supérieure 

Veiller à ce que la température de

à 40 °C (104 °F) 

fonctionnement soit comprise entre 10 °C

 

et 40 °C (50 °F et 104 °F).

Lorsque la température de surface est 

Température dépassant la plage de 

supérieure à 60°C (140 °F) 

mesure de température de 0° C à 60 °C 

 

(32 °F à 140 °F).

Summary of Contents for TRI TEMP

Page 1: ...Non contact thermometer INSTRUCTIONS FOR USE www trimedika com Part No CWC017 Revision 0720 EN...

Page 2: ......

Page 3: ...English...

Page 4: ......

Page 5: ...o take measurement of the patient X How to take measurement of an object XI Adjustment of settings XII Review previous measurements XIII Changing the batteries XIV Troubleshooting and Error messages X...

Page 6: ...the device This thermometer is designed for use by trained healthcare personnel only and must be used in accordance with the instructions in this manual All users must familiarise themselves with the...

Page 7: ...nergy Screen colour indicator which appears in BODY mode o GREEN 37 3 C 99 1 F Normal range o ORANGE 37 4 37 9 C 99 3 100 2 F Slight fever o RED 38 C 100 4 F Febrile Audible alarm if temperature is 38...

Page 8: ...ter to direct sunlight Do not autoclave the thermometer Do not point the thermometer at a heat source before use as it could raise the sensor temperature and produce erroneous results The thermometer...

Page 9: ...isposed of according to the institutional guidelines Patient Safety If the accuracy of the thermometer is in question use an alternative method to determine the patient s temperature and then ask your...

Page 10: ...n Fragile Keep Dry Serial Number IEC 60417 5034 Direct Current IEC 60417 5333 Type BF applied part Final TriTemp IFU Manual V3 0817 ENG D qxp_Layout 1 12 09 2017 12 24 Page 10 X IV Symbols WARNING Mee...

Page 11: ...t etc will affect the measurement 2 The sensor must be protected from dust liquids and other particles 3 The thermometer must be pointed directly at the forehead and NOT at an angle How does it work A...

Page 12: ...r passer au mode de lecture Le mot BODY temp rature corporelle clignote l cran o Appuyer sur le Bouton de lecture pour passer la temp rature de surface SURFACE TEMP o Appuyer sur le bouton de mise sou...

Page 13: ...pour le faire diminuer de 0 1 chaque fois Le d calage minimum possible est de 3 o Appuyer sur le bouton de mise sous tension 1 fois pour teindre l appareil 63 TriTempTM Non contact thermometer F TriTe...

Page 14: ...in 1 second Screen colour indicator appears o GREEN 37 3 C 99 8 F Normal range o ORANGE 37 4 37 9 C 100 100 2 F Slight fever o RED 38 C 100 4 F Febrile The device has a HIGH temperature alarm which w...

Page 15: ...user can the move through each MODE option from 1 to 4 as listed above by pressing the POWER button each time SETTING THE UNIT TEMPERATURE C OR F To change the unit of measurement from C to F o Press...

Page 16: ...READ button to switch audio ON o Press the POWER button 3 times to move through the MODE options and switch off the device SETTING THE READ MODE Body surface temp o BODY mode is the default setting an...

Page 17: ...used if a user wants to compare with another device e g mercury thermometer and this enables them to use a known offset value to give a comparable reading It is NOT recommended for any other applicati...

Page 18: ...k into place XIII Change the batteries TriTemp Non contact thermometer 16 XII Review previous measurements The TriTemp will retain measurements of the last 32 readings To access this information press...

Page 19: ...aking a temperature Air draft or rapid change in ambient temperature Ensure there is no direct air flow in the area of use Readings are too close together Ensure there is 3 5 seconds between temperatu...

Page 20: ...ng the lens except when cleaning is required The device exterior can be cleaned by wiping gently with a cloth moistened in a mild soapy detergent Ensure liquid does not enter the device and NEVER subm...

Page 21: ...tside those detailed in this document 3 Alteration or repair of the device not authorised by TriMedika 4 Accidental damage Shipping cost to return the device is not included in the warranty and a Retu...

Page 22: ...tteries Alkaline Measurement Units Celsius C or Fahrenheit F Operating temperature humidity 10 C 40 C 50 F 104 F 85 Relative Humidity Storage Conditions 20 C to 50 C 4 F to 122 F 85 Relative Humidity...

Page 23: ...setting is ON Consumption 150mW Auto Power OFF 30 seconds Temperature read time 1 second No of readings per device 40 000 measurements Ordering Details for Optional Accessories Product Number Name Des...

Page 24: ...meter EN 24 22 ENG Manufacturer s Declaration of the EUT Final TriTemp IFU Manual V3 0817 ENG D qxp_Layout 1 12 09 2017 12 24 Page 24 So i ii iii iv v vi vii viii ix x xi xii xiii xiv xv xvi xvi xvi x...

Page 25: ...riTemp TR1 Non contact Infrared Thermometer is intended for use in the electromagnetic environment specified below The customer or the user of TriTemp TR1 Non contact Infrared Thermometer should assur...

Page 26: ...or hospital environment Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode N A Mains power quality should be that of a typical commercial or hospital environment Final TriTemp IFU Manual V3 0...

Page 27: ...G Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 U T 95 dip in U T for 0 5 cycle 40 U T 60 dip in U T for 5 cycles 70 U T 30 dip in U T for 25 cyc...

Page 28: ...iTemp Non contact thermometer 46 XV Richtlinien TriTemp erf llt EU Richtlinie 93 42 EC bez glich Medizinischer Produkte TriTemp ist r ckverfolgbar zu ASTM E 1965 98 und Europ ischem Standard EN60601 1...

Page 29: ...rs such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To asses...

Page 30: ...XIII Batterien Wechseln Die mit dem Ger t gelieferten 2 x AAA Batterien erm glichen bis zu 20 000 Messungen Batteriesymbol Wenn der LCD Bildschirm das Batteriesymbol zeigt m ssen die Batterien ersetz...

Page 31: ...mitter where P is the maximum out put power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 M Hz the separation distance for the higher frequenc...

Page 32: ...vale Campus 398 Springfield Road Co Antrim Belfast Northern Ireland BT12 7DU United Kingdom Email Info trimedika com Tel 44 2890 265140 www trimedika com Contact us at Final TriTemp IFU Manual V3 0817...

Reviews: