107
FUNCTION SETTING USING SETTINGS
MEASURE (TEMPORARY SETTING)
MEASURE contains settings related to measurement functions that may be temporarily
changed.
Unlike initial setting, this function is used for performing measurements temporarily under cer-
tain conditions.
After power on or after clearing all measurement values, the system returns to the initial sta-
tus.
1
Tap MEASURE of the index of the "SETTTINGS screen."
Set items of "temporary setting" and the Set Condition button are displayed.
1/1
MONO
MANUAL
3
オフ
LEVEL3
‑
R-L Measure
Auto Mode
Cont. Cycle
IOL MODE
Target Brightness
Cyl Sign
MEASURE
INITIAL
DISPLAY
PRINT INT
PRINT EXT
COMM
NETWORK
SPECIAL
EXIT
SETTINGS
R-L MEASURE (CONTINUOUS MEASUREMENT OF RIGHT AND LEFT EYES)
Both-eye measurement (R-L) or single-eye measurement (R or L) can be set. Before ship-
ment, the default setting is both-eye measurement (R-L).
1
Tap the Set Condition button of the set item display "R-L Measure" of "MEASURE."
2
Tap the Setting Change button and select either "BINO" or "MONO."
3
Setting is done.
1/1
SETTINGS
BINO
MONO
MEASURE
INITIAL
DISPLAY
PRINT INT
PRINT EXT
COMM
NETWORK
SPECIAL
EXIT
R-L Measure
Auto Mode
Cont.Cycle
IOL Mode
Target Brightness
Cyl Sign
BINO
MANUAL
3
OFF
LEVEL 3
-