ALADDIN
- User manual
Rev. 19 EN of 10/07/2019
9
2 PRECAUTIONS
This electronic instrument is a precision tool and it is intended to be used in professional healthcare
environment including hospitals, physician’s offices, surgical centers and limited care facilities, where
equipment and systems are administered by healthcare professionals. Make sure to use it and keep it in a
suitable place, at a normal temperature, humidity and atmospheric pressure out of direct sunlight.
To ensure proper functioning, install the instrument in a vibration-free location.
Connect all cables correctly before use.
Use the recommended network voltage.
When the instrument is not in use, turn off the power supply and protect it from the sun and from
dust.
To obtain accurate and reliable measurements, keep the measuring cone clean and free of dust.
This product conforms to the EMC standard (IEC 60601-1-2 4th Edition).
ELECTRICAL MEDICAL DEVICES require special EMC precautions and must be installed and activated
in accordance with the EMC instructions provided in the accompanying documentation.
Use of accessories and cables other than those supplied with the instrument, except cables sold by
the equipment manufacturer as spare parts, may lead to an increase in emissions and reduce the
device's or system's immunity.
The eventual cables connected to USB and LAN ports must be less than 3 meters length.
The device should not be used adjacent to or stacked with other equipment; if adjacent or stacked
use is inevitable, the equipment should be observed to verify normal operation in the configuration
in which it will be used.
Portable RF communications equipment should be used no closer than 30 cm to any part of the
equipment, including specified cables. Otherwise, degradation of the performance of this equipment
could result.
Failure to follow warnings related to electromagnetic compatibility, can compromise the essential
performance or basic safety of the device, affect the proper operation of the camera and software,
false alarms, interruption of operations or wrong measurement.
In presence of high level EMC disturbance (i.e. Electrostatic Discharge, Electrical fast transient/Burst)
the device could stop working. In case, don’t use on patient until the source of disturbance is
removed.
Care is exercised in design and manufacturing to minimize damage to devices under normal use.
However, electronic devices are susceptible to many environmental stresses. The device may be
affected by electrostatic discharge (ESD). In an environment likely to cause ESD, such as air
conditioning, humidification, non-conductive floor coverings, synthetic clothing, discharge any
charge collected on your body before touching the device.
The FDA labelling for some IOLs contain sizing based upon white-to-white measurements derived from
studies in which this measurement is done with callipers. It is unknown whether the white-to-white
measurement from this device yields results systematically biased compared to those from calliper
measurements. Thus, sizing based upon white-to-white measurements from this device may not be
consistent with those based upon measurements with callipers.
Summary of Contents for ALADDIN HW3.0
Page 1: ...ALADDIN User manual Rev 19 EN of 10 07 2019 1 ALADDIN HW3 0 User manual Rev 19 EN 2019...
Page 27: ...ALADDIN User manual Rev 19 EN of 10 07 2019 27...
Page 36: ...ALADDIN User manual Rev 19 EN of 10 07 2019 36 Figure 14 Figure 15...
Page 44: ...ALADDIN User manual Rev 19 EN of 10 07 2019 44 Figure 22 Figure 23...
Page 91: ...ALADDIN User manual Rev 19 EN of 10 07 2019 91...