ALADDIN
- User manual
Rev. 19 EN of 10/07/2019
85
The first-step interface has quite the same structure as the spherical IOL calcualtion. The available toric lenses
you can select come from a list of models whose calculation constants and optical parameters have been
validated. The user can in case insert new toric manufacturers and/or models inside toric IOL settings section
(see 13.4.3).
In addition to choosing the “Target”, you need to specify also the “Surgical Induced Astigmatism (SIA)” and
“Incision Location (IL)”. The former identify the astigmatism (in diopters) induced by the incision while the
latter identify the surgical incision axis.
After having selected the toric IOL model, a values table from which the Spherical Equivalent Power is
obtained. Once you choose a lens, pressing “Next” at the bottom right, you enter in the second-step of toric
IOL calculation.