ALADDIN
- User manual
Rev. 19 EN of 10/07/2019
15
When operating the chinrest up/down switch, be careful not to pinch the patient’s hand. The
patient may be injured.
4.2 Electrical safety
To avoid risk of electric shock, this device must be connected to supply mains with protective
earth.
ALADDIN HW3.0 has an on-board power supply unit installed. For connection to the mains, use only the
manufacturer-approved cables provided with the device.
Before performing maintenance on the device, turn it off and disconnect the power cable.
Maintenance activities must be done in absence of the patient.
Fuse change can be done by the user by following the instructions and safety precautions described in
this Manual.
Do not touch the LAB/USB ports contacts and the patient at the same time.
4.3 LED emission safety
CAUTION - The light emitted from this device is potentially hazardous. The longer the duration of
exposure, the higher the risk of ocular damage. Exposure to the device light when using it at maximum
intensity will infringe the safety indication after a 60-minutes use.
ALADDIN HW3.0 has a series of LEDs of various types and powers installed. All the characteristics are detailed
in the Technical Specifications section in this manual.
The LED groups comply with the emission limits for the Group 2 instruments according to the standard ISO
15004-2.
4.4 Installation with external devices or IT Network
ALADDIN HW3.0 complies with the CE marking requirements.
Before connecting an external device, such as a computer, printer, monitor, keyboard, mouse or
other devices, make sure that they comply with the EN 60950-1 standard and have the CE marking.
An external keyboard or another input device compatible with “keyboard wedge interface” (PS/2) such
as barcode or card readers can be connected to the device to input text.
When ALADDIN HW3.0 is installed in rooms for medical use, the PC and the connected printer must be
powered by means of an IEC 60601-1 compliant insulating transformer.
If ALADDIN HW3.0 is installed in rooms for medical use without a computer, it is not necessary to use an
insulating transformer.
Do not use mobile phones or other devices not compliant with the requirements of class B EMC close to
ALADDIN HW3.0.
Every external device that has to be connected to ALADDIN HW3.0 must have a connection cable
(USB or LAN) with a maximum length of 3 m.
After connection of external devices to the USB or LAN, the end installator must check that the system
maintains basic safety and essential performance of the product in compliance with IEC 60601-1.
Summary of Contents for ALADDIN HW3.0
Page 1: ...ALADDIN User manual Rev 19 EN of 10 07 2019 1 ALADDIN HW3 0 User manual Rev 19 EN 2019...
Page 27: ...ALADDIN User manual Rev 19 EN of 10 07 2019 27...
Page 36: ...ALADDIN User manual Rev 19 EN of 10 07 2019 36 Figure 14 Figure 15...
Page 44: ...ALADDIN User manual Rev 19 EN of 10 07 2019 44 Figure 22 Figure 23...
Page 91: ...ALADDIN User manual Rev 19 EN of 10 07 2019 91...