
ALADDIN
- User manual
Rev. 19 EN of 10/07/2019
138
13.6 Admin
Figure 94
This is the instrument's administration panel (Figure 94).
It provides certain information on the system: serial number (S/N) and software version (S/V).
The
“Check”
button starts the calibration check procedure
.
Check the calibration
See the paragraph describing the procedure
It is absolutely essential to check the calibration when the device has been transported from
one place to another and when it has suffered an impact or thermal shocks.
It is recommended to check the measurements every day when turning on the device.
The “Enable RX/AL Trend Function” gives access to the function for reviewing biometry data over
time in charts from the exam list view. See section 11.16 for further details.
The “Application” frame manages the behaviour of the integrated software:
Upgrade
Updates the integrated software
Backup
Starts the backup procedure on a USB driver
Close App
Closes the application
Privacy
Refer to section 13.7
Summary of Contents for ALADDIN HW3.0
Page 1: ...ALADDIN User manual Rev 19 EN of 10 07 2019 1 ALADDIN HW3 0 User manual Rev 19 EN 2019...
Page 27: ...ALADDIN User manual Rev 19 EN of 10 07 2019 27...
Page 36: ...ALADDIN User manual Rev 19 EN of 10 07 2019 36 Figure 14 Figure 15...
Page 44: ...ALADDIN User manual Rev 19 EN of 10 07 2019 44 Figure 22 Figure 23...
Page 91: ...ALADDIN User manual Rev 19 EN of 10 07 2019 91...