11.2 Overview
Every working environment has the same screen layout. In Figure 10 a sample of the acquisition screen is
shown.
Figure 10
11.3 Checking the calibration
The calibration must be checked:
every day before starting patient examinations;
when the device has been transported from one place to another;
when it has suffered an impact or thermal shocks.
Right eye data
Left eye data
Live images from camera
The measurement selections are
enlightened in orange.
The active working area is enlightened
in orange.
Main menu
Settings