CORRECT alignment
WRONG alignment
Figure 13
Now check and several times acquire the calibration checking device using the complete acquisition (K-AL-
ANT). For details on this acquisition mode, see chapter 11.5.
If the calibration is ok, the “Valid” word will be display for all the measurements (Figure 14). If the
measurements are incorrect, the words “Repeat” or “Not Valid” will be displayed besides the wrong
measurement (Figure 15).
See table below for instructions.
MESSAGE
MEANING
WHAT TO DO
VALID
Good
calibration
Click on “Main” to start a new examination.
When asked, press Yes to save the “Calibration Check”.
NB: the “Valid” word must be displayed for ALL the measurements.
REPEAT
Acquisition
failed: cannot
check
calibration
Acquire again until three times following these suggestions:
improve the environmental conditions (less light and no
reflections on the sphere);
clean the sphere of the calibration tool;
make sure that the calibration tool is position correctly.
If the calibration check is still not valid, do not take any patient
measurement and contact Technical Support.
NOT
VALID
Incorrect
calibration