To check the calibration, turn on the device, and when asked to check the calibration, press START. The test
patient is automatically created.
Figure 11
The next screen explains how to check calibration
:
Figure 12
Set the calibration tool supplied with the device in the special holes in the chin rest and press until the tool
is blocked on the device. Check that the calibration tool is perfectly aligned with the device. If the calibration
tool is positioned correctly, all the rings of the Placido disk should be seen reflected in the center on the
surface of the hemisphere (Figure 13).