ALADDIN
- User manual
Rev. 19 EN of 10/07/2019
81
In the section “PRE-OP KERATOMETRY” you can insert the Keratometry data that was measured before the
first cataract surgery.
In the section “IMPLANTED IOL DATA” you can insert information about the IOL implanted in the first cataract
surgery:
Model: allows to insert the implanted IOL model:
o
selecting from the on-board database of Spheric IOL or Toric IOL lenses, by using the button
which opens the selection list
o
Inserting manually the description of the implanted IOL
A Constant/ LF: are the calculations constants used for the calculation performed to select the power
of the IOL implanted in the first cataract surgery, are filled automatically if the lens model is selected
from the on-board archive or must be inserted manually for manually inserted models.
Adjust the appropriate Lens Factor/A Constant for IOL by subtracting 0.25 mm from Lens Factor "bag"
constant if IOL in sulcus.
IOL Power (SEQ): is the spherical Equivalent power of the Implanted IOL
Toricity: is the cylinder value selection list for the toricity of the implanted IOL, it’s automatically
populated with a list of cylinder values and submodels if the implanted IOL information has been
selected from the on board database
Otherwise if the IOL model information has been inserted manually, this list is locked to “Manual”
and the cylinder value must be inserted manually
Summary of Contents for ALADDIN HW3.0
Page 1: ...ALADDIN User manual Rev 19 EN of 10 07 2019 1 ALADDIN HW3 0 User manual Rev 19 EN 2019...
Page 27: ...ALADDIN User manual Rev 19 EN of 10 07 2019 27...
Page 36: ...ALADDIN User manual Rev 19 EN of 10 07 2019 36 Figure 14 Figure 15...
Page 44: ...ALADDIN User manual Rev 19 EN of 10 07 2019 44 Figure 22 Figure 23...
Page 91: ...ALADDIN User manual Rev 19 EN of 10 07 2019 91...