[ How to create a noise table ]
1) Open Menu tab [6-1] and then set the noise table number (maximum10 points).
2) Open Menu tab [6-2], and set distance parameters into the data field in accordance
with selected numbers. (For example, if “10” is set at [6-1], you will need to set data
[600]~[609] for distance values.)
3) Set distance data for [600], and push “ENTER key” of PC to complete the entry.
Similarly set distance values for [601]…[609].
4) After completing input of distance values, the next step is entering amplitude values.
5) Open Menu tab [6-3] and set threshold level parameters into the data fields in
accordance with selected numbers. For example, if “10” is set at [6-1], you will need to
set data [610]~[619] for amplitude values.)
6) Amplitude points must correlate to the distances defined under [6-2]. Select table
address
(
ex. [610] and enter amplitude data, then push “ENTER key” of PC to
complete the entry. Similarly, set amplitude data for [611]…[619].
7) After setting all data, then click “WRITE”.
Setting of distance data must be in ascending order, that is, from shortest to longest
distance. Noise table error and noise table malfunction will occur if settings are mis-
ordered. Error status will appear under [7-3]
Example of a noise table is shown above. All noise lower than the noise threshold line will be ignored.
1
2
3
4
5
6
7
8
[60 0]
[601]
[602]
[603]
[604]
[605]
[606]
[607]
[610]
[611]
[612]
[613]
[614]
[615]
[616]
[6167
S
ignal
A
mpl
itude
Distance
Noise Table
Surface Echo
5- 3