[4--] Calibration
[4-1] 4mA Calibration
Parameter
Default Value
Description
3.50-4.50
4.00
In fixed current output mode with current set at
4mA, input ammeter reading.
[4-2] 20mA Calibration
Parameter
Default Value
Description
15.00-25.00
20.00
In fixed current output mode with current set at
20mA, input ammeter reading.
[4-3] Offset Calibration
Parameter
Default Value
Description
-1.00-1.00
0.00
Parameter range in metric [m] units
If measurement unit is [ft], display will be converted to show ‘feet’ units.
[4-4] Span Calibration
Parameter
Default Value
0.90-1.10
1.00
[5--] Advanced Setting
[5-1] Averaging Time
Parameter
Default Value
Description
1-120
10
Measuring value is averaged in accordance with
[sec] setting
[5-2] Noise Margin
Parameter
Default Value
Description
0-100.00
3.00
Noise is eliminated in accordance with [dB]
setting
When noise level is high, noise can be masked by attenuating the signal gain.