63
―
HXPRM10mnCT002E
(6)
Difference
Sum
Average
)
Caluclates the input of VOLT, TC, RTD or SCALE, and output.
Example)
When subtracting the Channel 1 input data from the Channel 2 input data to set the
difference (0 to 40 mV)
Records and indicates “Channel 2 (input value) minus Channel 1” to Channel 2
Hold down the
“
MENU
”
key for 3 seconds or more
to enter the setup mode. Use the
△
key to indicate
"
". Press the
“
ENT
”
key.
Use the
△
key to select the channel you want to set.
Press the
“
ENT
”
key.
Use the
△
key to select "
". Press the
“
ENT
”
key.
Use the
△
key to select the "channel you want to
subtract." Press the
“
ENT
”
key.
Use the
△
key to select a numeral. Use the
key
to shift a digit. Press the
“
ENT
”
key.
Use the
△
key to select a numeral. Use the
key
to shift a digit. Press the
“
ENT
”
key.
Press the
“
ENT
”
key. Setting is completed.
To return to the user mode, hold down the
“
MENU
”
key for 3 seconds or more.
About Difference, Sum, and Average
The channel to be set must be larger than the one on which computes the "difference, sum, or
average". Set up only Channel 2.
The range and scale of the set channel is the same as the channel on which computes the
"difference, sum, and average".
A range value after the computation cannot exceed the maximum range of the instrument.
Set for the reference channel only the voltage, current, thermocouple, resistance temperature
detector, and their scaling ranges.
Display
Operation Keys
Description
(
Reference channel
)
(
Setting channel
)
(
Span input value
after difference operation
)
(
Zero input value
after difference operation
)
MENU
ESC
3 sec
ENT
ENT
ENT
ENT
ENT
FEED
ENT
FEED
ENT
[
Note
]
An entry of an incorrect numeral indicates an error,
,
or
. Press the
“
ENT
”
key and re-enter a correct numeral.