115
―
HXPRM10mnCT002E
10.1.7 When the Chart Paper is not Fed at All
10.1.8 When the Chart Paper is not Fed Properly
Flow chart 6
No chart paper feed at all
Chart holder
attached
properly?
Contact our dealer where you
purchased the instrument, or our
sales representative.
YES
YES
NO
Set properly (See 5.1 on Page 25)
Chart paper fed by
FEED key?
DI contact for
remote operation
OFF?
Change Chart speed.
Turn on the DI contact
(See
7.2.9 on Page
89)
YES
NO
NO
Flow chart 7
Chart paper feed improper
Chart cover
locked properly?
Chart paper set
properly?
Chart guide
attached
properly?
Attach properly
(See 5.1 on Page 25)
NO
NO
NO
YES
YES
YES
Attach properly
(See 5.1 on Page 25)
Used in adequate
environment?
Attach properly
(See 5.1 on Page 25)
NO
YES
Contact our dealer where you
purchased the instrument, or our
sales representative.
Secure an adequate
environment
(See 11.1.5 on Page 119)