20
―
HXPRM10mnCT002E
Fig. 4.5 Input Wiring ( RTD, mA )
CAUTION
①
Attach the shunt resistor to the input terminal block of the instrument.
②
Input accuracy is effected with the shunt resistor. Use the following recommended resistor.
Resistance
:
250
Ω
Rated power
:
1/4W Tolerance
:±
0.1% max.
Temperature coefficient
:±
50ppm max.
+
-
B B A
+
-
B B A
B B
A
Shunt registor
+
-
①
RTD input
②
mA input