Model: TTT-710
6.2 Various Conditions Setting
- 67 -
•
Potentiometric titrator (TP-70)
•
Automatic burette (ABT-511
)
(c) When this product is connected to the following analytical instruments, the conditions set here
become invalid and the conditions set at the analytical instrument become valid.
>>
Instruction manual of analytical instruments
•
X Series analytical instruments
•
R Series analytical instruments by “RS-232C” connection
•
Automatic titrator
Setting Procedure
Operation Screen
example
①
To the Conditions Setting Item Selection screen
⋅⋅⋅
Press
SET
at the Measurement Waiting screen.
Condition Setting Item
Selection screen
②
To the Measurement Condition Setting screen
⋅⋅⋅
Press
or
and move the cursor (blinking) to “2.
Meas. Condition” and press
ENTER
.
③
Setting
⋅⋅⋅
Set the measurement conditions as follows:
: Move the cursor (blinking) of the setting item value
(last digit).
: Change the value.
ENTER
: Enters the set value.
•
After setting, the display returns to the Condition Setting
Item Selection screen.
Cursor (blinking)
Measurement Condition
Setting screen
④
End the setting.
⋅⋅⋅
Press
SET
.
•
The display returns to the Measurement Waiting screen.
(3) RS-232C communication conditions setting
(a) When the product is operated by analytical instrument and RS-232C control, or when PC is
connected and commands are sent or received by RS-232C, the communication conditions must be
set. Select the one of the setting range.
(b) When the product is used by connecting analytical instruments, set the communication conditions
as shown in the following table:
RS-232C Communication Conditions Setting Items
Communication
condition item
display
Communication
condition name
Setting range
Set value when
connecting to analytical
instrument
Baud Rate
Baud rate
9600, 4800, 2400, 1200, 600, 300 bps
4800 bps
Data Length
Data length
7 Bit, 8 Bit
8 Bit
Parity
Parity
NONE, ODD, EVEN
NONE
Stop Bit
Stop bit
1 Bit, 2 Bit
2 Bit