(1) Specifications
Product name
Turntable
Model TTT-710
Display
20 digits x 4 lines, character LCD display
Number of table specimens
12, 18, 36, 60, 100 specimens (100 specimens table is optional)
Usage beaker size
12 specimens
200mL beaker, 300mL tall beaker
(Our recommended product: AGC TECHNO GLASS or HARIO glass beaker)
18 specimens
100mL tall beaker
(Our recommended product: AGC TECHNO GLASS or HARIO glass beaker)
36 specimens
30mL beaker, 50mL tall beaker
(Our recommended product: AGC TECHNO GLASS glass beaker)
60 specimens
20mL dedicated beaker (NICHIDENRIKA GLASS Model H-20)
100 specimens
20mL dedicated beaker (NICHIDENRIKA GLASS Model H-20)
Stirring method
Standard: Magnetic stirrer method
When options connected: Propeller stirring method
External control input/output
RS-232C (control for Automatic titrator AUT-701/AUT-501, X Series
HM-42X/CM-42X/MM-43X, R Series HM-30R/CM-30R/MM-60R,
dedicated titrator TA-70/TS-70/TP-70)
OPTION 2 (R Series, G Series) control input/output
Automatic burette (ABT-511) / Flow analysis reference unit (FAR-201A)
control input/output
External control box control input/output
Air pump box control output
Liquid level sensor signal input (washing water 1, waste solution)
Power output for propeller stirring unit
Washing method
Standard: Pure water shower washing
When options attached: Pure water bubbling washing (air pump box (with
bubbling) when waste water valve attached)
Liquid chemical shower washing (when external
signal box, reagent pump, reagent washing reservoir
attached)
Liquid chemical bubbling washing (when external
signal box, reagent pump, reagent washing reservoir,
air pump box (with bubbling), waste water valve
attached)
Pure water soak washing (when waste water valve
attached)
Liquid chemical soak washing (when external signal
box, reagent pump, reagent washing reservoir, waste
water valve attached)
Air blow (when air pump box attached)
Washing sequences
8 kinds and user washing
Washing time
0 to 9999 seconds for all washing modes
Number of air blows
1 to 9 times
Number of repetitive measurements
1 to 9 times
Measurement end detection
Detected by end detection pin or end table No. setting