BP209 Series
6 The Graphics User Interface (GUI)
Rev: 8.1, 14-Feb-2022
© 2022 Thorlabs
Page 21
Reset Application Settings to Defaults
This button resets all BEAM application settings to defaults:
Application Setting
Default Value Explanation
Gain Control
Automatic
Gain is set to automatic.
Bandwidth Control
125 kHz
Instrument set to continuous data acquisition.
Baseline Correction
Automatic
The ADC output signal that is caused by environ-
mental and stray light is compensated.
Aperture Width
9.0 mm
Full Aperture Width is used for calculations.
Active Slit Pair
5 µm
Slit Pair that is used for measurement
Scanning Mode
Slit
Normal mode, in contrast to Knife Edge Mode
Target Scan Rate
10 Hz
Scan Rate set value.
Scan Rate Correction
Enabled
Enabling Scan Rate Correction, a control loop will
maintain the scan rate close to the target value.
Calculation Area Preset Auto
Rectangle
Software determines a rectangular area in which the
power level is above the clip level.
Calculation Area Clip
Level
1.0 %
For defining the Calculation Area in automatic mode,
the lower intensities limit is set to 1% of the peak in-
tensity to clip noise.
Averaging Mode
None
No averaging
Clip Level
13.50 %
The clip level for calculating the ellipse and beam
width is set to 13.5% of the difference between the
baseline and the peak intensity.
Hold Maximum
Disabled
After reaching the max. number of data to be plotted,
with the next data acquisition the first data set will be
discarded and the the most recent set will be added.
Autoscale to Peak
Enabled
The diagrams that show intensities, are scaled to the
peak intensity.
Correct Beam Width
Enabled
Measurements results are beam-width corrected.
Reference Position
Sensor
Center
Reference position is set to the aperture center. Cen-
troid and peak positions are calculated relative to the
reference position.
Power Unit
mW
All power units are displayed in mW.
Plot Method
Latest Data
After reaching the max. number of data to be plotted,
with the next data acquisition the first data set will be
discarded and the the most recent set will be added.
Plot Interval
Every Meas.
For every measurement, the data are plotted
Max. Data Points
100000
The maximum number of data to be plotted
Velocity
200 mm/s
Translation speed of the stage
Language
The language of the application can be selected.
Summary of Contents for M2MS
Page 2: ...Copyright 2022 Thorlabs Version Date 8 1 14 Feb 2022 ...
Page 73: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 71 Example ...
Page 92: ...BP209 Series 7 Operation Instructions Page 90 21424 D02 ...
Page 119: ...BP209 Series 7 Operation Instructions Rev 8 1 14 Feb 2022 2022 Thorlabs Page 117 ...
Page 158: ...BP209 Series 11 Appendix Page 156 21424 D02 11 10 Drawing M2MS BP209 ...
Page 162: ...BP209 Series 11 Appendix Page 160 21424 D02 11 14 Certifications and Compliances ...
Page 167: ...www thorlabs com ...