BP209 Series
7 Operation Instructions
Rev: 8.1, 14-Feb-2022
© 2022 Thorlabs
Page 75
Preparation
·
Set Gain Mode to "
Automatic
".
·
Set X and Y Slit Bandwidth to "
1000 kHz
".
·
Set Base Line Mode to "
Manual
".
·
Set Photodiode Bias to "
On
" - this shortens the
rise/fall time of the photo diode.
·
Set Aperture Width to "
0.9 mm
" - the software
uses only an area of Ø 0.9 mm in the sensor center
for calculation.
·
Set Active Slit Pair to "
25 µm
".
·
Set Scanning Mode (for alignment) to "
Slit
".
·
Set Scan Rate to "
2 Hz
".
·
Enable Scan Rate Correction.
Enable X and Y profiles.
Disable Gaussian and Bessel fits and zoom in the profiles. A proper alignment is reflected in a
maximum amplitude and a flat top of the profile. The flat top is a result of the small beam size