6
•
Operation
Vanquish Charged Aerosol Detectors Operating Manual
Page 101
Parameters
Description
Gas regulator pressure
The default unit setting for the gas pressure corresponds
to the
Pressure Unit
that is defined on the detector
configuration pages in the Chromeleon software.
Gas regulator
pressure nominal
Specifies the nebulizer gas pressure as specified on the
certificate that was shipped with the nebulizer.
Observe the notes in
section 6.5.2 Verifying the Nebulizer
Gas regulator
pressure
Displays the actual pressure of the gas supplied to the
nebulizer.
After you have set the nebulizer gas pressure, monitor the
GasRegPressure
parameter until it reaches the setpoint.
Gas regulation mode
Sets a specific nebulizer gas pressure for the analysis,
based on the gas pressure specified for the nebulizer from
the nebulizer certificate.
As a standard, the
Analytical
mode is set.
For VH-detector only:
Two modes for regulating the nebulizer gas pressure in the
detector are available:
Analytical
mode and
MicroLC
mode.
For details on the gas regulation modes, see
section 6.7.6
Peak width
Defines the width of a chromatographic peak at the half of
the peak height. Specify a peak width for the narrowest
expected peak in your chromatogram.
As a standard, when you set the
Peak Width
, Chromeleon
automatically calculates values for the
Data Collection
Rate
and the
Filter Constant
.
Data collection rate
Defines the number of data points per second (Hz) that
the Chromeleon software collects from the detector and
stores as raw data.
For further information, see
section 6.7.3 Data Collection
Rate and Filter Constant, page 104
Filter constant
Applies a filter constant to the output current to
electronically reduce the baseline noise during the
analysis.
For further information, see
section 6.7.3 Data Collection
Summary of Contents for Vanquish VF-D20
Page 4: ...Contacting Us Page 4 Vanquish Charged Aerosol Detectors Operating Manual...
Page 28: ...2 Safety Page 28 Vanquish Charged Aerosol Detectors Operating Manual...
Page 38: ...3 Detector Overview Page 38 Vanquish Charged Aerosol Detectors Operating Manual...
Page 44: ...4 Unpacking and Transport Page 44 Vanquish Charged Aerosol Detectors Operating Manual...
Page 150: ...7 Maintenance and Service Page 150 Vanquish Charged Aerosol Detectors Operating Manual...
Page 177: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 177 12 Index...
Page 181: ...12 Index Vanquish Charged Aerosol Detectors Operating Manual Page 181...